Medium Priority Alerts
The following medium priority alerts are accompanied by a triple beep, repeated every 30 seconds,
and a yellow light flashing at 0.5 Hz.
Message
Breathing not detected
Check cannula
Low oxygen
Remove battery
Cold battery
Failed pressure sensor
Failed piping
Reminder: If you are not in close proximity to the Horizon® P5, you may not be able to hear
or see medium priority alerts. Make sure the Horizon® P5 is in a location where alerts and
alarms will be recognized if they occur.
30
Condition / Action / Explanation
The concentrator has not detected a breath for 15 seconds. Check that
the cannula is connected to the concentrator, that there are no bends
in the tube and that the cannula is correctly positioned in your nose.
Oxygen concentration has been between 60% and 82% for 10 minutes.
If the situation persists, switch to your backup oxygen source and
contact your service provider.
The battery has exceeded the temperature limit and charging has
stopped. Connect the device to an external power source and remove
the battery to cool it down. If possible, move the concentrator to a
cooler location. If the condition persists, contact your service provider.
The battery temperature is below 0°C. Move the concentrator to a
warmer location and wait for the device to reach recommended
operating temperatures before turning it on.
Check your filters for blockages. If the pressure sensor is defective or
out of order, an alarm will be triggered by the device. Contact your
service provider.
Nozzle failure (hoses or control valves). Contact your service provider.
DOMU722_v7_EN_20221006_Horizon® P5 User manual