Information about the text sections
Messages may be displayed in this section. They are intended to guide the user on the status of this
device. The following section is devoted to presenting these messages and explaining their meaning.
When two conditions occur simultaneously, the condition with the higher priority will be
displayed.
Notifications & alerts
Information Messages
The following information is NOT accompanied by an audible alarm or visual change in the indicators:
Message
Startup in progress
Please wait
Setting X Battery
XX%
Audible notifications, alert the user of problems. In order for the auditory notifications to be heard,
the maximum distance the user can move away from the notification should be determined based
on the ambient noise level.
The Horizon® P5 monitors various parameters during operation and uses an intelligent alarm system
to signal a malfunction of the concentrator. Mathematical algorithms and time delays are used to
reduce the likelihood of false alarms while ensuring proper notification of an alarm condition.
If a low or medium level alarm is triggered, the buzzer will emit two or three sounds depending on
the priority level, the LED next to the "alarm" button will flash or be steady, the alarm logo and the
error message will appear on the display.
DOMU722_v7_EN_20221006_Horizon® P5 User manual
Condition / Action / Explanation
Displayed when the power button has been pressed for two seconds.
The concentrator is starting the system.
"X" represents the selected setpoint setting. "XX%" represents the
approximate percentage remaining on the battery charge.