Do not connect the Horizon® P5 oxygen concentrator in parallel or in series with other
oxygen therapy devices.
Wind or strong drafts can adversely affect the accuracy of the delivery of oxygen therapy.
Some breathing efforts of the patient may not trigger the concentrator and thus send a
bolus. It is recommended to have a stress test with your doctor before using Horizon® P5.
Do not leave the HORIZON P5 connected to the power supply without its battery.
Never leave the Horizon® P5 in an environment which may reach high temperatures. It
could damage the device.
In the event of a device failure, contact your authorized service provider or the
manufacturer's service center immediately. Do not disassemble or repair it yourself.
In order to get the best performance from the oxygen concentrator, SCALEO Medical
recommends using the oxygen concentrator for more than 30 minutes each time.
Frequent use of the oxygen concentrator for a short time may shorten the life of the
device.
Avoid the use of extension cords.
Do not drop or insert foreign objects into any openings. It is strictly forbidden to block
the air inlet and exhaust port of this device or to place the device on a soft surface, such
as a sofa or bed, which may cause the block the exhaust. Keep the air inlet away from
lint, hair, or objects alike.