Hub/Router Connection; Firmware And Communications Troubleshooting; Table 4-5. Troubleshooting - Firmware And Communications - Thermo Scientific Mercury Freedom 80i Manual

Table of Contents

Advertisement

Troubleshooting
System Troubleshooting Guides

Hub/Router Connection

Hub/Router Connection
Firmware and
Communications
Troubleshooting
4-12
Mercury System Manual
When connecting instruments within the Mercury System or connecting a
local PC to the Mercury System via a hub or router, use straight cables. Do
not use the UPLINK port unless you are connecting to a network or
another hub.
When connecting instruments within the Mercury System or connecting a
with UPLINK
local PC to the Mercury System via a hub or router with an UPLINK port,
use a crossover cable on the UPLINK port connection and a straight cable
on the other port connections.
Table 4–5 describes possible firmware and communications malfunctions
and provides the recommended action to take to restore the system to
normal operation.
Table 4–5. Troubleshooting – Firmware and Communications
Malfunction
80i does not
communicate with 81i or
82i
80i does not
communicate with 81i
81i will not respond to
Gas mode commands
from the 80i
Possible Cause
80i was powered up before
81i/82i were powered up
Units are not connected via
secure cables
81i is in Service mode
80i configured improperly
81i in Service mode
Incorrect Ethernet cables
used
Ethernet cables
disconnected or
malfunctioning
Ethernet hub malfunction
Action
Shut down the three instruments.
Power up 81i and 82i, wait 60
seconds, then power up the 80i.
Check cable connections.
Verify 81i is not in Service mode,
set to OFF, if necessary.
Check 81i IP address under
Diagnostics > System
Configuration and ensure it
reflects actual 81i IP address, left
justified.
Verify 81i is not in Service mode,
set to OFF, if necessary.
When attempting to
communicate directly between
instruments, a crossover cable is
required. If using an Ethernet
hub/switch port (other than
UPLINK), straight cables are
required.
Verify cable functionality, replace
if necessary.
Verify hub LEDs are lit.
Thermo Fisher Scientific