Thermo Scientific 3962 Operating Manual page 22

Environmental chamber
Table of Contents

Advertisement

| Product Description
Chapter 5
|
5-6
Model 3962
Figure 5-8
Connectors on Board
4. Route the wires through the auxiliary hole located on the back of the unit. See
Figure 5-8 on page 5-6 for the location of this hole and Figure 5-9 on page 5-6 for
routing details.
Figure 5-9
Board Location in Drawer
5. When wiring is completed, re-install the unit top.
NOTE
Accuracy of the analog outputs measured at the circuit board terminal strip
with respect to the environmental chamber display is ±1 unit. There is no
calibration of the analog outputs by the environmental chamber, so the
instrument(s) connected to the analog outputs must be calibrated to match the
environmental chamber display before returning to service.