Thermo Scientific 3962 Operating Manual page 41

Environmental chamber
Table of Contents

Advertisement

Thermo Scientific
The glass door may be cleaned with commercial glass cleaner or with a solution
of ammonia and water.
Health hazard
The surfaces of the workspace may be contaminated. Contact with contaminated
cleaning liquids may cause infections. Disinfectants may contain harmful
substances.
When cleaning and disinfecting, always observe the safety instructions and
hygiene regulations!
Wear safety gloves.
Wear safety goggles.
Wear mouth and respiratory system protection gear to protect your mucous
membranes.
Observe the safety instructions of the disinfectant's manufacturer and the
hygiene supervisor.
Incompatible cleaners
Some device components are made of plastic. Solvents can dissolve plastics.
Strong acids or caustic solutions can cause to become brittle of the plastic. For
cleaning plastic components and surfaces, do not use solvents that contain
hydrocarbons, solvents with an alcohol content of more than10% or strong acids
or caustic solutions!
Decontamination or cleaning agents
The "Thermo or Forma" company or their agent is to be consulted if there is any
doubt about the compatibility of decontamination or cleaning agents.
Cleaning and Disinfection |
Chapter 8
Model 3962 | 8-3