Operational Safety Rules; Warranty - Thermo Scientific 3962 Operating Manual

Environmental chamber
Table of Contents

Advertisement

| Safety Notes
Chapter 1
|
1-2
Model 3962
Drying time after decommissioning.
Humidity After transport or storing under humid conditions a drying-out process
must be performed. During the drying-out process the equipment cannot be
assumed to meet all the safety requirements of the IEC 61010-2-010 standard.
The drying-out period is 2 hours.
Should you encounter problems that are not detailed adequately in these operating
instructions, please contact Thermo Electron LED GmbH immediately for your own.

Operational Safety Rules

The following rules must be heeded when working with environmental chambers:
Observe the sample weight limits specified for your environmental chambers, whole
and its shelving in particular; see "Specifications" on page 11-1.
Arrange the samples evenly throughout the work space, making sure not to place
them too closely to the interior walls to ensure a uniform temperature distribution.
Do not load your environmental chambers with substances that exceed the
capabilities of the available lab apparatus and Personal Protection Equipment to
provide sufficient degrees of protection to users and third parties.
Check the door seal every 12 months for proper sealing performance and possible
damage.
Do not process any samples containing hazardous chemical substances that may be
released into the ambient air through defective seals or may cause corrosion or other
defects on parts of the environmental chamber.
The user is responsible for carrying out appropriate decontamination procedures
when hazardous materials are spilled on or inside the environmental chamber.
If the environmental chamber is not used in the manner specified in this operating
manual, the protection provided by the equipment design may be impaired.

Warranty

Thermo Fisher Scientific warrants the operational safety and functions of the
environmental chambers only under the condition that:
The environmental chamber is operated and serviced exclusively in accordance with
its intended purpose and as described in these operating instructions.
The environmental chamber is not modified.
Only original spare parts and accessories that have been approved by Thermo
Scientific are used (third-party spares without Thermo Scientific approval void the
limited warranty).
Inspections and maintenance are performed at the specified intervals.
An operation verification test is performed after each repair activity.