Totalizing Application
s
Preparation
§ Turn scale on: Press the e key
§ While all segments are lit, press the
) key
§ Select Setup: Press the k key
repeatedly until appl is displayed
§ Confirm appl: Press the ) key
§ Select the Totalizing application:
Press the k key repeatedly and
confirm with the ) key
Totalizing Application Parameters
3� 6�
Minimum load for autosave
3�6�1*
1 digit
3�6�2
2 digits
3�6�3
5 digits
3�6�4
10 digits
3�6�5
20 digits
3�6�6
50 digits
3�6�7
100 digits
3�6�8
200 digits
3�6�9
500 digits
3�6�10
1000 digits
3�16� Autosave mode
3�16�1*
Off
3�16�2
On
3�17� Individual/Component printout
when saved
3�17�1
Automatic printing off
3�17�2*
Print the entire standard
print configuration
every time with the
O key
3�22� Source of values for data saved
automatically
3�22�1*
Application 1
3�22�2
Application 2
3�23� Saved value
3�23�1*
Net
3�23�2
Calculated
3�23�3
Net and Calculated
* = Factory setting
§ Save the settings with the ) key and
exit Setup: Press the ( key several
times�
68
Signum 3 Ex
Minimum Load
The minimum load required for
automatic taring of the container
weight on the platform (first weight)
is configured in Setup under:
Appl
:
m.weigh
:
3.5.
The minimum amount that a
component must weigh before it
can be saved in totalizing memory
is configured in Setup under:
:
appl 3
:
totaliz
3.6.
You can choose from the following
10 levels for this setting:
1 digit
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
The "digits" here refer to the scale
intervals in the connected weighing
platform� If the interval of the
connected platform is 1 g, for example,
and 1000 digits are required, you must
place at least 1000 g (= 1000 intervals)
on the weighing platform for autotaring
(only with the "Autotare first weight"
option selected)�
Printout
In Setup menu:
:
appl 3
:
totaliz
3.17.
You can configure whether a printout
is generated manually by pressing
the p key or automatically when a
weight value is stored in the totalizing
memory�
If you select the 3.17.1 menu item,
printouts can only be generated
manually by pressing the p key
(single printout)�
If the 3.17. 2. menu item is selected
(Print one component), the component
record is printed�
The total data record is printed when
you clear the totalizing memory
(by pressing the c key)�