Summary of Contents for Minebea Intec Combics Series

  • Page 1 Presented By Dallas Ft Worth Austin Houston NicolScales.com 800.225.8181 Contact Us Nicol Scales & Measurement is an ISO Accredited Calibration Company that has provided calibration, repair and sales of all types of weighing and measurement products since 1931.
  • Page 2: Service Manual

    Service Manual Minebea Intec Combics Series Models CAW... | IP69K | IP44 Combics Complete Scales, Models CAH... | CAS... Combics High-Resolution Complete Scales, Models CAI... | IP69K | IP44 Combics Complete Indicators, Models CAAP... Combics Weighing Platforms WCA5001-e18038...
  • Page 3: Table Of Contents

    Contents Operating Design ............13 Table of Contents Activating the service mode ............13 Service Mode ................13 Service Manual ..............1 CAIS1, CAISL1, Activating the service mode for Combics Introductory Notes ............4 CAIS2, CAISL2 ............13/14 Regarding this Manual .................4 Additional Menus in the Service Mode .........
  • Page 4 Contents Adjustment data .................. 90 Repairs for Combics Indicators ........58 Determining the Resistance for Adjustment Front Plate ..................59 of the Off-center Load ..............96 Power Supply PCB ................60 Corner load filling ................98 Measurement Points ................61 change index ..............100 Power Cord ....................
  • Page 5: Introductory Notes

    Introductory Notes Regarding this Manual This Service Manual provides information on maintenance, calibration and error messages for Combics series devices. If you would like more information or would like to make a comment, please send an e-mail to Int.Service@Minebea-intec.com Symbols Used...
  • Page 6: Safety Instructions

    These Combics devices conform to relevant guidelines and standards. However, improper use of devices can result in mechanical or electrical damage. Repairs should only be carried out by factory-trained service technicians. Work on Combics series devices requires special experience. For this reason, absolutely no unauthorized repairs should be attempted! Warning! Disconnect the equipment from the power supply before starting repair work or replacing components.
  • Page 7: General Information

    The Combics series is an upgraded version of the Combics series from Minebea Intec. The newly developed digital PCB improves the compatibility of connection options for the series. Before working on the Combics series, we advise that you attend a service course offered by Minebea Intec. General Notes For device care and maintenance, see User manual CAIS/CAISL1,..2| Publication no.: WCA6007-d10051...
  • Page 8: Overview Of Models

    Overview of Models Minebea Intec Combics indicator units (individual delivery) or complete scales (scales with an indicator unit) are mainly used in industrial applications. Due to the various options available, these models are built in accordance with customer wishes. The series consists of different platforms from 3 kg to 3000 kg in non-stainless steel or stainless steel with corresponding resolution options.
  • Page 9 Introductory Notes | General Model Description Protection class CAWIS1 Complete scale IP69K CAWIS2 Complete scale IP69K CAWIS3 Complete scale IP69K CAWISL1 Complete scale IP44K CAWISL2 Complete scale IP44K CAWISL3 Complete scale IP44K CAH1E/3E | CAS... Complete scale Indicator IP69K Weighing platform IP65 CAH1G/3G Complete scale Indicator IP69K Weighing platform IP65...
  • Page 10: Auxiliary Tools

    Introductory Notes | General Auxiliary Tools Warning! After removing the seal marks (warranty sticker) and concluding the service work, please seal Combics devices with your company sticker. Please note that in addition to the correct tools, a stable and clean work space is required, which must be free of vibrations and air drafts.
  • Page 11: Special Tools

    Introductory Notes | General Special Tools Quantity Description Insulating transformer AC/DC multimeter RS 232/485 interface converter DMS simulator Test connector for COM port test (COM 1) Socket wrench set (size 5, 5.5, 7) SartoCas H8S Service Manual Combics...
  • Page 12: Handling And Operation

    Handling and Operation Combics Key Function CAIS1, CAISL1, CAIS2, CAISL2, CAIS3, CAISL3 Detailed information about indicator key functions is contained in the respective operating instructions. CAIS1, CAISL1 Display, On/Off key. General function keys: Zero, Tare, Switch function (e.g. Gross/Net), Adjustment/Calibration, Print/Data output. CAIS12, CAISL2 Display, On/Off key, General function keys:...
  • Page 13: Cais3, Caisl3

    Handling and Operation CAIS3, CAISL3 Display (graphic-capable dot-matrix display), On/Off key. General function keys: (Scale) When connecting a second weighing platform, the display switches between the platforms, Zero, Tare, Switch function, Adjustment/Calibration, Print/Data output, (x10) Toggle unit between normal and 10-fold higher resolution, (B/G) Net/Gross value, Clear function, Reference value, Apply value, Switch display, Toggle between available applications,...
  • Page 14: Operating Design

    Operating Design Activating the service mode Service Mode The service mode enables access to an expanded setup menu. You must be in this mode to carry out the most important calibration and adjustment work for the indicator and for the connected weighing platform(s).
  • Page 15 Operating Design | Service Mode Combics1 | 2 Important! The Service password can be entered via the decimal keypad on Combics 2 series indicators. disp_blin_d.eps For Combics 1: t Use the k key (or p key) to select 2 and confirm with the ) key, the cursor moves to the second digit, enter the next number.
  • Page 16: Additional Menus In The Service Mode

    Operating Design | Service Mode Combics1 | 2 Additional Menus in the Service Mode In the Text menu (Setup): The following are the menu items displayed behind the Date (DATE) and Password (U-CODE) menu items – Service date (S-DATE) – Serial number of the terminal (SER.NO) –...
  • Page 17: Activating The Service Mode For Combics Cais3, Caisl3

    Operating Design | Service Mode Combics3 Activating the Service mode for Combics CAIS3, CAISL3 t Switch on Combics (e key). y The scale is now in an application program (i.e. "Weighing" application, the example shows the "Counting" application). t Enter the service password via the decimal keypad and confirm with the M key. Important! The number shown in the figure is not identical to the service password.
  • Page 18 Operating Design | Service Mode Combics3 Exiting the Service Mode t Switch the Combics 3 off and then on. WP 1 WP 2 y The device is now in the normal Application mode. COM 1 COM 2 I/O control If you exit the Setup menu without changing settings by confirming with the M key or Bar code Config.
  • Page 19 Operating Design | Service Mode Combics3 The CAL key function and Activate ext. adj. sub menus of the Service Mode Overview “Calibration/Adjustment” menu also contain additional menu When activated, the Service mode enables service technicians items. access to additional menu items in the Setup menu which are If one of the COM 1 or COM 2 interfaces has been configured not displayed when the Service mode is not active.
  • Page 20 Operating Design | Service Mhhode Combics3 Configuring the A/D converter Combics | 1 | 2 Open the menu access switch t Remove the cap that covers the menu access switch on the back of the indicator t Slide the switch to the left (= "open position") Enter the service password as described above (202122), confirm with the ) key and scroll back using the ( key.
  • Page 21 Operating Design | Service Mode Combics 1 | 2 ADC configuration y Select the ADC Configuration menu item and confirm: PARAM.! t Press the k key several times, ADC-CON " appears, press the ) key. until " STAND y Select whether a standard configuration " "...
  • Page 22 Operating Design | Service Mode Combics 1 | 2 ADC configuration Verifiable configuration " VERIF. "RANGE 2" " "RANGE 3" Class: CLASS "MAX." " " " " Unit (select the unit) UNIT " " IIII “ " e.g.: kilogram or ton “...
  • Page 23 Operating Design | Service Mode Combics | 1 | 2 ADC configuration Close the menu access switch. Slide the menu access switch to the right ("closed position"). ADC Quick Configuration for Combics 1 and 2 There is a quick installation option available when connecting a weighing platform (WP 1 or WP2) to a Combics 1 or 2 Open the menu access switch t Remove the cap that covers the menu access switch on the back of the indicator...
  • Page 24 Operating Design | Service Mode Combics | 1 | 2 ADC configuration ADC-CON " appears on the display y " ADC CON S-CODE and then " " S CODE t Enter the service password as described previously (202122). t Press the ) key. y The device is now in Service mode.
  • Page 25 Operating Design | Service Mode Combics 3 ADC configuration Configuring the A/D converter Combics 3 Open the menu access switch t Remove the cap that covers the menu access switch on the back left of the indicator t Slide the switch to the left (= "open position") Switch on the Combics 3 y The scale is now in an application program (i.e.
  • Page 26 Operating Design | Service Mode Combics 3 ADC configuration y The letter "S " is displayed in the Setup menu header. The device is now in Service mode. y The Setup menu has been expanded along with the User menu. To view or change device parameters in Setup mode: t Select the Device parameters"...
  • Page 27 Operating Design | Service Mode Combics 3 ADC configuration Important! If " WP 2 " is selected with the soft key and confirmed with the > soft key, RS232 " " will appear first in the display. Confirm once more here with the > soft key.
  • Page 28 Operating Design | Service Mode Combics 3 ADC configuration INTERNAL y " " appears in the display. tConfirm using the > soft key. ADU.CON y " " appears in the display. tConfirm using the > soft key. y " Standard "...
  • Page 29 Operating Design | Service Mode Combics 3 ADC configuration Exiting the Service Mode t Switch the Combics 3 off and then on. The device is now in the normal Application mode. If you exit the Setup menu without changing settings by confirming with the M key or soft key, the Service mode will remain active.
  • Page 30 Operating Design | XBPI Protocol Settings XBPI Protocol Settings The XBPI protocol settings are made so that adjustment routines can be carried out using the SARTOCAS or Psion service software without having to activate the BPI mode via the SBI BPI keypad. Note: This menu item can be accessed in the standard menu.
  • Page 31: Test Program For Combics Indicators| 1 And | 2

    Operating Design | Test Program Combics 1 | 2 Test Program for Combics Indicators| 1 and | 2 A test program is included on the main/digital PCB of the Combics indicator. When this test is run, the action being carried out or the action to be carried out by the user is shown on the display.
  • Page 32 Operating Design | Test Program Combics 1 | 2 – The test program automatically detects whether or not an optional interface is available for E5 or L9 and runs the corresponding test routine. Important! There are cable options for which not all interface lines are used. These are not checked by the test program.
  • Page 33: Test Preparation

    Operating Design | Test Program Combics 1 | 2 Test Preparation The test adapter must be connected to the COM 1 interface before running the internal test routine. If the UNICOM interface is used, a second adapter must also be connected to it. Important! However, this does not apply to the B9 (Ethernet), B1 (profibus) and A3 (RS-485) options.
  • Page 34 Operating Design | Test Program Combics 1 | 2 t "DISPLAY " appears on the display. t The display back lighting turns off and then on again. y All segments in the display are activated. y Carry out the visual inspection. t All LEDs turn off and then on in sequence (for Combics 2).
  • Page 35 Operating Design | Test Program Combics 1 | 2 y The option test takes place. If option L9 is available, the battery symbol appears at the end of the test. For option A10 (ADU 1x 10.000e), the scale symbol appears. y The test takes place for the standard COM 1 interface STD UART "UART STD"...
  • Page 36: Test Results

    Operating Design | Test Program Combics 1 | 2 Important! If beep tones sound (short and long) during the test, this means that the test program has discovered an error. Pay attention to the test results "RESULT." TEST OK y Once the test is complete, "RESULT" appears on the display. If the test finished without any errors, "TEST OK"...
  • Page 37: Test Program For Combics Indicator 3

    Operating Design | Test Program Combics 3 Test Program for Combics Indicator 3 A test program is included on the digital PCB of the Combics indicator. When this test is run, the action being carried out or the action to be carried out by the user is shown on the display.
  • Page 38: Test Implementation

    Operating Design | Test Program Combics 3 Important! However, this does not apply to the B9 (Ethernet), B1 (profibus) and A3 (RS-485) options. These options do not require a test adapter. The test adapter must be supplied with a mains adapter (15V=) for option A5. Test Implementation t Switch Combics to Standby mode.
  • Page 39 Operating Design | Test Program Combics 3 Important! You can speed up the test program by pressing the t key. t All LEDs turn off and then on in sequence (for Combics 3). Then they all turn on and then off together. y "KEYBOARD TEST"...
  • Page 40: Test Results

    Operating Design | Test Program Combics 3 Test Results If the test program detected an error, then this is also shown at the end of the test on the display. If the program discovers multiple errors, then these are all shown on the display contrary to Combics 2.
  • Page 41: Error Description

    Operating Design | Error Description Test Combics | 1 | 2 | 3 Error description After completing the internal test, the individual test items are displayed with their respective results. A successful test is confirmed with the message "Test OK." An attempt is always made to test options such as E5 or A10. If these options are not available, then the test results would say NOT FOUND.
  • Page 42 -Digital PCB defective not all segments are displayed -Display unit defective No message -Cable not connected or connected Operating Design | Error Description Test Combics | 1 | 2 | 3 LEDs do not light as described in incorrectly to display unit Fehler! Verweisquelle konnte nicht -Cable to display unit defective gefunden werden.
  • Page 43 Operating Design | Error Description Test Combics | 1 | 2 | 3 The following error messages are possible: SCL2-RS485 LOOPBACK If A7 is available: -Option A10 (not allowed with A7) No data connection via -Connection cable between digital PCB and the IS interface interface not inserted or defective -A7 interface defective...
  • Page 44: Test Adapter For Combics 1| 2 | 3

    Operating Design | Test Adapter Test Combics | 1 | 2 | 3 Test Adapter for Combics 1| 2 | 3 Test adapters are required for testing interfaces. They connect the output signals of the interfaces with their inputs. Different adapters are required depending on the device (CAIS or CAISL) and whether or not CAIS devices have a cable with a DSUB25 or a Mini-DIN connection.
  • Page 45 Operating Design | Error Description Test Combics | 1 | 2 | 3 If the device is equipped with a digital PCB with screw terminals, then the electrical connections must be established at the screw terminals of the PCBs. The connections are listed in the following table: Test Adapter for Option A6/A62 ( RS-232)
  • Page 46 Operating Design | Error Description Test Combics | 1 | 2 | 3 Test Adapter for Option A9 ( 4-20mA/0-10V) Adapter for interface option A9 (4-20mA / 0-10V). Because an automatic test is not possible for this option, LED indicators are used. The wiring of the adapter is included in the following figure.
  • Page 47 Operating Design | Error Description Display Combics | 1 | 2 | 3 Error messages in the display Error messages are shown in the main display. Error messages are shown permanently; Information messages are shown for 2 seconds. After this the program automatically returns to the weighing mode. Service Manual Combics...
  • Page 48 Adjustment could not be completed Allow to warm up again and repeat the inf 03 within a certain time adjustment process Built-in calibration weight defective Contact your local Minebea Intec Service Center Inf 06 Display Cause Solution Service Manual Combics...
  • Page 49 Operating Design | Error Description Display Combics | 1 | 2 | 3 inf 07 Function not allowed in scales not verified Contact your local Service Center for information for legal use on changing settings inf 08 The load on the scale is too heavy to enable Check whether in your configuration “zero at Inf 09 zero positioning...
  • Page 50 Operating Design | Error Description Display Combics | 1 | 2 | 3 Data not found or unreadable Contact your local Service Center inf 73 Function is blocked None inf 74 (e.g., menu is locked) No weighing platform connected Connect a weighing platform inf 98 No weighing platform connected Connect a weighing platform...
  • Page 51: The Combics 1 Indicator Includes

    Device Description Device Description Minebea Intec Combics indicators or complete scales are mainly used in industrial applications. Due to the various options available, these models are built in accordance with customer wishes. The Combics 1 indicator includes: – Front plate with keypad and digital display –...
  • Page 52: Options For Cais1, Caisl1 Indicators

    Device Description Options for CAIS1, CAISL1 indicators Main PCB for IP44 version with D-SUB plug. UNICOM: RS-232: Option A1 RS-422: Option A2 RS-485: Option A3 Opt. A1-A3 /A5,A9 Digital I/O: Option A5 4-20mA: Option A9 ALIBI MEMORY: Opt. B1/B9 Alibi memory: Option E5 COM1 Main PCB for IP69 version...
  • Page 53: The Combics 2 Indicator Includes

    Device Description The Combics 2 indicator includes: – Front plate with keypad and digital display – Combics indicators (CAIS2 or CAISL2) come with either a D-Sub 25 (IP44) or cable glands (IP69). Unicom interfaces in several versions com1 interface in 2 versions PS-2 interface (possible in two versions) ADC, option A8...
  • Page 54: Options For Cais2, Caisl2 Indicators

    Device Description Options for CAIS2, CAISL2 indicators COM1 Digital PCB for IP44 version with D-SUB plug. PS/2 UNICOM: RS-232: Option A1 RS-422: Option A2 RS-485: Option A3 Opt. Opt. B1/B9 A1-A3 / A5, A9 Digital I/O: Option A5 4-20mA: Option A9 ALIBI MEMORY: Alibi memory: Option E5...
  • Page 55: The Combics 3 Indicator Includes

    Device Description The Combics 3 indicator includes: – Front plate with keypad and digital display – Combics indicators (CAIS3 or CAISL3) come with either a D-Sub 25 (IP44) or cable glands (IP69). Unicom interfaces in several versions com1 interface in 2 versions PS-2 interface 2*3000e multi-interval...
  • Page 56: Options For Cais3, Caisl3 Indicators

    Device Description Options for CAIS3, CAISL3 indicators COM1 Digital PCB for IP44 version with D-SUB plug. PS/2 UNICOM: RS-232: Option A1 RS-422: Option A2 RS-485: Option A3 Opt. Opt. B1/B9 A1-A3 / A5, A9 Digital I/O: Option A5 4-20mA: Option A9 ALIBI MEMORY: Alibi memory: Option E5...
  • Page 57: Options For Combics Pcbs In Combination With The Following Options

    Device Description Options for Combics PCBs in combination with the following options CAISL1 (ADC 2x 3.000e) ( RS232) ( RS485) 7-65901-210-00 main board C1 (Onboard ADC) 7-65901-200-00 main board C2/3 7-65901-260-00 interface board ADC/RS485 7-65901-220-00 ADC-board 10.000e CAISL2/3 (ADC 2x 3.000e) ( RS232) ( RS485) 7-65901-200-00 main board C2/3...
  • Page 58 Device Description Overview of options for Combics indicators UNICOM: RS-232: Option A1 RS-422: Option A2 RS-485: Option A3 Digital I/O: Option A5 4-20mA: Option A9 Ethernet: Option B9 Profibus: Option B1 ALIBI MEMORY: Alibi memory: Option E5 Theft protection Option L7 L8 (24V) INDUSTRIAL POWER SUPPLY: 24V.
  • Page 59: Repairs For Combics Indicators

    Repairs for Combics Indicators Opening Combics Indicators Important note! An insulating transformer must be installed upstream when working on an opened Combics indicator under power. t Ten nuts (see left image) must be removed to open the Combics indicator. Replacing the Seal Important! After carrying out any repair, the seal between the front plate and the housing must be checked visually for damage and replaced if necessary.
  • Page 60: Front Plate

    Repairs | Component replacement | Front Plate Replacing the Front Plate If the foil-covered keypad, the keys or the display is defective, the front plate must be completely replaced. There are three front plate versions: The Combics 1 indicator (left) includes: ST 3 Front plate with keypad and digital display The Combics 2 indicator (left) includes:...
  • Page 61: Power Supply Pcb

    Repairs | Component Replacement | Power Supply PCB Display blank If the display is blank, the power supply to the device or display may have been interrupted. t Disconnect the device from the power, open it and pull the plugs from the components. t Then connect the device to the power via an insulating transformer and then reconnect all components in succession and measure the...
  • Page 62: Measurement Points

    Repairs | Component Replacement | Measurement Points Main PCB Measuring Points Note: The voltages are measured using a digital voltmeter (DC setting) against the ground. Because there are no special measurement points, the checks must be made with the greatest care using a thin test probe (danger of a short circuit).
  • Page 63 Repairs | Component Replacement | Measurement Points Digital PCB measuring points M 1: 15 V ±1.5 V input voltage for the DC/DC converter M 2: +3.3 V ± 0.2 V +5 V ± 0.2 V M 4: +8 V ± 0.5 V M 5: -8 V ±...
  • Page 64: Power Cord

    Repairs | Component Replacement | Power Cord Replacing the Power Supply Connection Cable If 15 V is not being supplied to the main digital PCB, then check the power supply PCB as already described. The device power supply may be interrupted. t Disconnect the device from the power and open it t Then replace the power supply cable.
  • Page 65 Repairs | Component Replacement | Main Digital PCB Replacing the Main/Digital PCB If the main/digital PCB is defective, then it must be replaced. Before replacing the digital PCB, you should attempt to read the converter controller using the H8S service software SartoCas H8S to enable you to load the old calibration/ adjustment data from the old weighing system to the controller of the new AD converter after replacement.
  • Page 66 Repairs | Component Replacement | Main Digital PCB Main/Digital PCB Pin Assignment COM1 Screw terminals CAIS(L) Connection Function ST4a LOAD PRINTER ST4a RESET_OUT 5V_OUT 5V switched n.c. LINE_OUT CTS_COM1 COM1 DTR_COM1 COM1 RXD_COM1 TXD_COM1 ST4a ST4a Universal In MINOR (<) PARES (=) MAJOR (>) Important!
  • Page 67 Repairs | Component Replacement | Main Digital PCB Screw terminals CAIS(L) COM 2/3 COM1 Connection Function COM1 LOAD PRINTER RESET_OUT 5V_OUT 5V switched Housing (ground) LINE_OUT CTS_COM1 DTR_COM1 RXD_COM1 TXD_COM1 Universal In COM1 MINOR (<) PARES (=) MAJOR (>) 5V switched PS2_Data PS2_Timer CTS_COM2...
  • Page 68 PCB interface description ADC Option (A8) Pin Assignment DSUB or screw terminals Combics 1 (on board, see left image) 2x3000e A8, 2x3000e A8, 2x3000e Connection Function EXC+ SENSE+ OUT+ OUT- SENSE- EXC- DSUB or screw terminal Combics 2 /3 (see right image) 2x3000e Connection Function EXC+...
  • Page 69 Interface Description | PCBs ADC Option (A10/A20) Pin Assignment DSUB or screw terminals Combics 1 (on board, see left image) 1x10.000e Connection Function EXC+ A20, 1x10.000e A10, 1x10.000e SENSE+ OUT+ OUT- SENSE- EXC- DSUB or screw terminals Combics 2/3 (see right image) 1x10.000e Connection Function...
  • Page 70 Interface Description | PCBs RS-232/485 Pin Assignment Option (A6/A7) DSUB or screw terminals Combics 1 interface RS-232/485 (see left image) A6/7, RS232/485 A6/7, RS232/485 DSUB or screw terminals Combics 2/3 (see right image) RS-232/485 interface Connection Function 1CTS Adjust Lock TxD/RxD+ TxD/RxD- LINE_OUT...
  • Page 71 Interface Description | PCBs RS-232/485 Pin Assignment Option (A62/A72) DSUB or screw terminals Combics 2/3 (see left image) Second RS-232/485 interface A62/72,RS232/485 Connection Function 1CTS Adjust Lock TxD/RxD+ TxD/RxD- LINE_OUT LINE_OUT A62/72,RS232/485 Service Manual Combics...
  • Page 72: Unicom Pcb Interface Description

    Unicom PCB Interface Description The interface options for the UNICOM interfaces are described here. Several of the interfaces can be configured via a switch. These interfaces must be configured before setting up the device. Important! The factory default setting for Opt.
  • Page 73 Interface Description | Unicom PCBs RS-232/RS-485 Option (A2/3) Pin Assignment Slot for UNICOM RS-422/RS-485 Interface The operating mode setting is carried out via switch SW1-4. The PCB factory setting is RS-485 operation (option A3). The switch must be reset to enable the card to operate via RS-422 (option A2). The other switches are used to activate the transmission side.
  • Page 74 Interface Description | Unicom PCBs Switch Function when switch is closed SW1-1 Load resistor, TX side 120 Ohm SW1-2 BIAS resistor, TX side (TXD+, Pull-Up) 680 Ohm SW1-3 BIAS resistor, TX side (TXD-, Pull-Down) 680 Ohm SW1-4 RS-422 operation= A2 (open: RS-485 operation = A3) SW2-1 Load resistor, RX side 120 Ohm SW2-2...
  • Page 75 nterface Description | Unicom PCBs Connection Function OUT5_LOW OUT5_HIGH OUT4_LOW OUT4_HIGH OUT3_LOW OUT3_HIGH OUT2_LOW OUT2_HIGH OUT1_LOW OUT1_HIGH IN5_LOW IN5_HIGH IN4_LOW IN4_HIGH IN3_LOW IN3_HIGH IN2_LOW IN2_HIGH IN1_LOW IN1_HIGH Service Manual Combics...
  • Page 76 Interface Description | Unicom PCBs 4-20mA / 0-10V interface, Option (A9) Slot for UNICOM 4-20mA /0-10V Interface. The operating mode setting is carried out via switches SW1-1 and SW1-2. The PCB factory setting is 4-20mA operation. The switch must be reset to enable the card to operate via 0-10V.
  • Page 77 Interface Description | Unicom PCBs Profibus Option (B1) Interface Slot for UNICOM interface. BIAS activated Different interface PCBs are used here for the various device types (CAIS and CAISL). For the Completion activated IP44 version, the PCB is used together with the adapter PCB. The adapter PCB is inserted into the UNICOM interface slot and connected to the PCB via a foil cable.
  • Page 78 Interface Description | Unicom PCBs Ethernet Option (B9) Interface Different interface PCBs are used here for the CAT cable connection various device types (CAIS and CAISL). For the IP69K IP44 version, the PCB is used together with the adapter PCB. The adapter PCB is inserted into the UNICOM interface slot and connected to the PCB Opt.
  • Page 79 Interface Description Version with the junction box (5) Part no. 57004-428-01 The illustration on the left shows GF frame 650x500 Load cell 1 Load cell 2 Load cell 3 Load cell 4 Junction box Junction Box 1 Service Manual Combics...
  • Page 80 Interface Description Junction box; open. 1.1 Cable to load cell 1 2.1 Cable to load cell 2 3.1 Cable to load cell 3 4.1 Cable to load cell 4 5.1 Connecting cable for A/D converter / indicator Contacts to connect the cable 5.1 Pin Assignments in the Junction Box 1 Service Manual Combics...
  • Page 81 To connect the cable, the wires must be connected to the inner terminals (marked in the illustration on page 19). Note: The color coding of the connecting cable (A/D converter / indicator to junction box) might be different when connecting a non-Minebea Intec platform to the Midrics indicator. Service Manual Combics...
  • Page 82 Service Manual Combics...
  • Page 83: Devicenet Interface

    - Self-adhesive aluminum foil - ferrite core – „Fieldbus“ software interface description – EDS file: see Minebea Intec website „minebea-intec Software download: Select product: Fieldbus files“ Assembly Connect interface module YDO0…-xxx (DeviceNet interface) directly to the digital board of the scale, without configuration.
  • Page 84 DeviceNet Interface DeviceNet-Installation YDO02C-DN (B3) Service Manual Combics...
  • Page 85 DeviceNet Interface DeviceNet - Installation: - Gehäuse öffnen - Montieren sie das DeviceNet-Board (* 1) an das Mainboard (* 3) Service Manual Combics mit Abstandsbolzen und Schrauben...
  • Page 86 DeviceNet Interface DeviceNet - Installation: l l a - Open the housing - Mount the DeviceNet Board (* 1) to the motherboard (* 3) ö f ä u with bolts and screws i e r - Assemble the cable gland - Prepare the DeviceNet - Cable (2 *) f ü...
  • Page 87: Expanded View Diagram 1

    Expnaded Views Expanded View Diagram 1 Refer to the spare parts list in ISSS. Front plate Front plate complete for Combics 1 Front plate complete for Combics 2 Front plate complete for Combics 3 Seal Tag holder Service Manual Combics...
  • Page 88: Expanded View Diagram 2

    Expnaded Views Expanded View Diagram 2 Refer to the spare parts list in ISSS. PCBs for Combics 201 Main PCB (on board ADU 2 x 3000e), For Combics 1, DSUB, IP44 202 Main PCB (on board ADU 2 x 3000e), For Combics 1, clamping connectors, IP69K 203 Digital PCB (without ADC), DSUB, IP44 For Combics 1, 2, 3 with options:...
  • Page 89: Lower Part

    Expnaded Views Expanded View Diagram 3 Lower part Small parts set Power cord Service Manual Combics...
  • Page 90: Adjustment Data

    Adjustment data Adjustment data Eccentricity (EN45501 3.6.2 ,8.2 A.4.7) Test loads The test load, approximately a third of maximum capacity, has to be placed successivly over the center of the areas shown in the The tests loads used for the tests are shown in the test report. picture.
  • Page 91: Adjustment Data

    Adjustment data (Ecklast) Ablesbarkeit in g Justierung Wiederholbarkeit Richtigkeitsprüfung Außermittige Belastung nicht nicht eichfähig eichfähig 15000d 30000d Justier- Zulässige Zulässige Weig- Zulässige Model gewicht Prüflast Prüflast Toleranz Prüflast Toleranz bereich Toleranz (±) (±) 0,0005 CAW1..P 0,0005 CAW1::S 1 kg 3 kg 0,0015 1 kg 0,001 kg...
  • Page 92 60,0 0,03 0,025 CAW1..P 25,0 0,025 Adjustment data CAW1::S 50 kg 50 kg 0,050 50 kg 0,050 kg 50,0 0,05 CAAP..P 100,0 0,05 CAAP..S 150,0 0,075 0,05 CAW1..P 50,0 0,05 CAW1::S 100 kg 100 kg 0,100 100 kg 0,100 kg 100,0 0,10 CAAP..P...
  • Page 93 Ablesbarkeit in g Justierung Adjustment data Ein- Zwei- Zwei- Drei- Ein- Bereich Bereich Teilung Bereich Bereich Spezifikationen: Justier- 3000e 2x3000e 2x3000e 3x3000e 6000e Wäge- siehe EN 45501; Modell gewicht bereich -BCE -NCE -MCE -UCE -RCE umgesetzt in Sartorius- Software "SSS-M". CAW1..P CAW1::S 0,5 | 1...
  • Page 94 CAW1..P CAW1::S Adjustment data 10 | 20 10 | 20 5 | 10 | 20 20 kg CAAP..P CAAP..S CAW1..P CAW1::S 20 | 50 20 | 50 10 | 20 | 50 50 kg CAAP..P CAAP..S CAW1..P CAW1::S 50 | 100 50 | 100 20 | 50 | 100 100 kg...
  • Page 95 Adjustment data Modell Reada- Repea-ta- Off-center Linearity Preload load bility weight bility loading test range 1/2/3/4 in kg error (±) CAH-1-E / 3-E 16ED-H 16 kg 0.1 g 10 kg-F1 10 kg-0.2 g 5kg / 0.3g 4 / 7 / 12 / 19 / 0,2g 4 kg CAS...
  • Page 96 Load Cell Connection (Color Assignments) Adjustment data Bridge supply voltage Test voltage Brückenspeisespannung Messspannung Load cell Cable construction +Excitation/ -Excitation/ +Signal/ -Signal/ manufacturer +Supply/ -Supply/ +Ausgang -Ausgang +Sense -Sense Wägezellen Anmerkung zur +Eingang -Eingang +Output -Output Hersteller Kabelausführung d 5mm, 6 wires, green, shield grounded Sartorius Hamburg blue...
  • Page 97: Determining The Resistance For Adjustment Of The Off-Center Load

    Adjustment data Determining the Resistance for Adjustment of the Off-center Load Determining Resistance Values - Place the test weight on each load cell in turn and write down each displayed value. - Using the lowest displayed value as the basis, calculated the resistance values for the other load cells in accordance with the following equation: Equation: R = R(0) x G(D) / G(T)
  • Page 98 Adjustment data Sample Calculation Based on Diagram 2 Initial resistance in the load cell: 1000 ohms. - Test weight: 500 kg - Readout for reference load cell = 498 kg (load cell with the lowest readout value). - Readout for load cell being adjusted = 501.1 kg - G(D) = 2.3 kg R = R(0) x G(D) R = 1000 ohms x 2.3 kg...
  • Page 99: Corner Load Filling

    Adjustment data Corner load filing for single-cell Combics 60.000d platforms and scales (CAAPP1-...-S, CAW.P1-...-S) In order to reduce weighing platform deviations for off-center loads (cornerloads), strain-gage load cell can be filed down. Identifying corner load errors is routine for trained service technicians and is essen- tial in ensuring that corner load correction is professionally implemented.
  • Page 100 Adjustment data Service Manual Combics...
  • Page 101: Change Index

    change index Date change index Reason for change 07.03.2015 Inserted DeviceNet interface 27.09.2016 Expansion to CAS ... 24.03.2017 Change of name 27.03.2018 „Corner load filing for single-cell Combics 60.000d platforms and scales“ added Service Manual Combics...
  • Page 102 Service Manual Combics...
  • Page 103 No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Minebea Intec. All rights reserved by Minebea Intec in accordance with copyright law. The information and figures contained in these instructions correspond to the version date specified below.

Table of Contents