Operating Instructions Compact Scale Puro® - Basic 98628-000-61 Edition 1.0.0 02/18/2020 Checked on the basis of Puro® scale with software version 2.89 Minebea Intec Bovenden GmbH & Co. KG, Leinetal 2, 37120 Bovenden, Germany Phone: +49.551.309.83.0 Fax: +49.551.309.83.190...
Page 2
Any information in this document is subject to change without notice and does not represent a commitment on the part of Minebea Intec unless legally prescribed. This product should only be operated/installed by trained and qualified personnel. In correspondence concerning this product, the type, name, and release number/serial number as well as all license numbers relating to the product have to be cited.
5.3.1 Procedure on Setting up the Totalization Application................14 5.3.2 (Totalization) Displayed Values........................15 5.3.3 Viewing and Clearing Statistical Data......................16 Menu Settings............................ 19 Menu Mode................................19 Menu Navigation ..............................20 6.2.1 Menu Selection [APPLIC]..........................20 6.2.2 Menu Selection [METRO]..........................21 Minebea Intec EN-1...
WARNING indicates that death or severe, irreversible injury may occur if appropriate safety measures are not observed. Take the corresponding safety precautions. CAUTION Warning of personal injury. CAUTION indicates that minor, reversible injury may occur if appropriate safety measures are not observed. Take the corresponding safety precautions. Minebea Intec EN-3...
Page 6
Warning of damage to property and/or the environment. NOTICE indicates that damage to property and/or the environment may occur if appropriate safety measures are not observed. Take the corresponding safety precautions. Note: User tips, useful information, and notes. EN-4 Minebea Intec...
If there are grounds for a complaint, this must be brought to the attention of the delivery company immediately. A Minebea Intec sales or service point must be informed. Visit our website http://www.puroscales.com or contact your dealer.
Level the scale using the four leveling feet. Ensure that all leveling feet make contact with the floor. Turn the feet until the air bubble is centered in the level indicator. Each of the leveling feet must support an equal load. EN-6 Minebea Intec...
71 to 100 % Remaining Note: When battery symbol blinks fast, approximately 30 minutes working time is left. When [lo.bat] is displayed, the scale will shut off. Charging the scale must be performed in a dry environment. Minebea Intec EN-7...
Battery is to be replaced only by an authorized Puro® service dealer. Dispose of battery according to local laws and regulations. Connect printer A printer can be connected to the printer port on the underside of the device. EN-8 Minebea Intec...
Operating elements, see Chapter 4.1.1.3. 4.1.1.2 Display elements LCD Display Item Description Item Description Busy Auto Taring Symbol Plus symbol Data transfer Minus symbol Weight unit Zero-setting Battery charge Weighing Printer Icon Totalization NET, Preset Tare, Gross Minebea Intec EN-9...
Page 12
Aborts the cali- items. play and advances to bration in pro- Decrement the the next available gress. digit. setting. Advances to Advances to the next the previous menu or item. digit in input Increment the digit. process. EN-10 Minebea Intec...
Page 13
4 Device description Compact Scale Puro® - Basic 4.1.1.4 Rear Display The models LargeTall (LT) and SmallTall (ST) have an additional rear display, which shows the same display elements as the front display. (See Chapter 4.1.1.2). Minebea Intec EN-11...
Applications "Totalization", "Auto Tare" and "Automatic Print" can be activated via Menu. [OP.FUNC] [A.TARE] Auto Tare [TOT.SET] Totalization [PRINT] [A.PRINT] Automatic Print Application Programs 5.2.1 Weighing Mode Display the weight of the sample in lb, oz, lb:oz, kg or g. EN-12 Minebea Intec...
Page 15
2. Place container on the scale (in this example, 0.598 kg). 5.2.1.1 Stability If a weighing value is stable, the unit symbol is displayed. Stable weighing value: Unstable weighing value: Stable gross weighing value below zero (no unit visible): Minebea Intec EN-13...
Statistical data (total accumulated weight, min/max weights and total number of weighing objects) is stored in memory for review and printing. Totalizing works in combination with each application mode. Totalization manual is set to ON by default. 5.3.1 Procedure on Setting up the Totalization Application EN-14 Minebea Intec...
Unit (Yes) button. 7. Leave the setup menu with the button. Menu 5.3.2 (Totalization) Displayed Values The Totalization Mode is active when icon is displayed. Put the first item on the scale. Minebea Intec EN-15...
4. Put another sample on scale and follow same operation (Automatic or Manual mode). icon will keep flashing until the weight is removed. 5.3.3 Viewing and Clearing Statistical Data The pan must be empty to view the stored statistics. EN-16 Minebea Intec...
Page 19
Menu The scale will display the statistical information in following order: Number of weighing done (N = 7): Totalized value (TOTAL = 5.225g) Minimum value (Min = 0.47kg): Maximum value (Max = 1.485kg) Minebea Intec EN-17...
Page 20
Gross loads and net loads cannot be added to the same total. If the first load is a gross weight, following loads must also be gross weights. If the first load is a net weight, following loads must also be net weights. EN-18 Minebea Intec...
Mode Unit previous menu item. The 2 menu item [METRO] shown on display. When viewing the setting (in this example [APPLIC] - [WEIGH] - [STAB.RA] value 0.5d). Minebea Intec EN-19...
Leave menu Note: Some modes / units may not be available in all models. 6.2.1 Menu Selection [APPLIC] Enter this menu to select application to be used. Only activated units will be accessible with the (Mode) Button. Mode EN-20 Minebea Intec...
— END Leave menu level 6.2.3 Menu Selection [UNIT] Enter this menu to customize weighing unit to be used. Default settings are identified by an "*" UNIT — kg Kilogram — OFF disabled — ON enabled* Minebea Intec EN-21...
— 8 SEC Backlight off after 8 minutes of no activity — D.LIGHT Display Backlight — OFF disabled — ON enabled — AUTO turns on when a button is pressed or the displayed weight changes* — COM.EQU Communication Module EN-22 Minebea Intec...
— FOOTER Footer info — OFF disabled* — ON enabled — MODE Mode info — OFF disabled* — ON enabled — INFO has no effect in this device — OFF disabled* — ON enabled — TOTAL Total Value Minebea Intec EN-23...
Enter this menu to define PC Output parameters. Default settings are identified by an "*" PC.OUT — MODE PC Output Mode — OFF disabled* — MAN.OUT Manual output — MAN.STA Manual output — INT.OUT Interval output — AUT.OUT Automatic output EN-24 Minebea Intec...
Menu Selection [CAL.ADJ] Enter this menu to perform calibration and adjustment of your scale (see chapter 7). CAL.ADJ — CAL Initiates a span calibration procedure (zero and span) — LIN Initiates a linearity calibration procedure (zero, mid-point and span) Minebea Intec EN-25...
— OFF Unlock* — ON Locked — S.CAL.AD Calibration / Adjustment Menu — OFF Unlock* — ON Locked — RESET Restore factory setting of current menu — OFF Unlock* — ON Locked — END Leave menu level EN-26 Minebea Intec...
Unit Mode menu item. 3. Change menu item until [CAL.ADJ] is displayed. 4. Press (Yes) button to enter into sub-menu item. Sub-menu item [CAL] (Calibration) is displayed. Minebea Intec EN-27...
Page 30
8. For changing the calibration weight value, press the (No) button to edit the Unit value. The first digit is blinking [_015.000 kg]. 9. Press (Yes) button to accept and highlight the next digit. The 2nd digit is blinking [0_15.000 kg]. EN-28 Minebea Intec...
[--C--] will be displayed while calibration weight is stored. In case of any mistake during the calibration process [CALE] (Calibration Error) is displayed. 14. Remove the weight after finish. The scale is ready for weighing. [LIN] Linearity Calibration Procedure Enter into Menu Mode: Minebea Intec EN-29...
Page 32
5. Change menu item until [LIN] shows on display. 6. Press (Yes) button to start zero point calibration. [0 kg] shows on display. [0] is blinking. 7. Empty the weighing pan. 8. Press the (Yes) button to confirm. EN-30 Minebea Intec...
Page 33
[00150.0] (in this example, 150 kg). 11. Put 2 linearization weight on the pan. 12. Press (Yes) button to start linearity adjustment at 100% of scale capacity. [--C--] is displayed while reading is stored. Minebea Intec EN-31...
(Settings from 0 … 31 are available). Refer to table to determine the GEO factor that corresponds to your location. Enter into Menu Mode: Press and hold until [M.E.N.U] displayed. Menu Release the button. The 1 menu item [APPLIC] shown on display. EN-32 Minebea Intec...
Page 35
(Yes) button to return to sub-menu selections. 11. Press (No) button to return to the first item in the current menu. Unit 12. Press the button to exit the setup menu and return to the weighing mode. Menu Minebea Intec EN-33...
There are some special response commands. These Commands are used as Standard responses. For example, error or confirmation. Special Response Commands always have the same size of 5 Bytes! The Scale confirmed the command execution without any errors. EN-36 Minebea Intec...
Page 39
8 SBI Interface Compact Scale Puro® - Basic ERROR The Scale has reported an Error during command execution. LOCKED The command can not be executed because the parameter is currently locked. Minebea Intec EN-37...
WARNING Improper repairs can pose considerable risks to the user. Repairs should only be performed by Minebea Intec authorized dealers using original spare parts. Cleaning 9.2.1 Cleaning instructions The device should be cleaned at regular intervals.
Before disposing of or scrapping the product, any batteries should be removed and taken to a suitable collection point. Please see our T&Cs for further information. We reserve the right not to accept products that are contaminated with hazardous substances (ABC contamination) for repair. Minebea Intec EN-39...
Compact Scale Puro® - Basic 11 Troubleshooting 11 Troubleshooting The Table lists common problems and possible causes and remedies. If the problem persists, contact Minebea Intec or your authorized dealer. Symptom Possible Cause Remedy Cannot turn on No power to scale...
Off, 0.5, 1 or 3 divisions Safe overload capa- 150 % of rated scale capacity city Levelling compo- Externally visible level indicator and adjustable, non-slip levelling feet nents Power = 5 V, 2A, AC adapter or portable rechargeable lithium battery Minebea Intec EN-41...
Page 44
(w x d x h) Platform size (w × l) 218 x 180 mm Shipment dimension 370 x 370 x 220 mm mm (w x d x h) Net weight 2.6 kg Shipping weight 4.1 kg EN-42 Minebea Intec...
Accumulation line If Printx → Content → Total is All or Result MIN: 11.11 kg Accumulation line If Printx → Content → Total is All MAX: 14.85 kg Accumulation line If Printx → Content → Total is All EN-54 Minebea Intec...
Reorient or relocate the receiving antenna. Increase the separation between the equipment and receiver. Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. Consult the dealer or an experienced radio/TV technician for help. Minebea Intec EN-55...
Page 58
Published by Minebea Intec Bovenden GmbH & Co. KG | Leinetal 2 | 37120 Bovenden, Germany Phone: +49.551.309.83.0 | Email: info@minebea-intec.com www.minebea-intec.com...