Averaging (Animal Weighing) Application
Averaging (Animal
Weighing) Application V
With the Averaging application, you
can use your weighing platform for
calculating weights as the average
of a number of individual weighing
operations� These individual operations
are also known as "subweighing
operations� "
This function is used to determine
weights under unstable ambient
conditions or for weighing unstable
samples (such as live animals)�
Characteristics
– Averaging started manually or auto-
matically� Configured in Setup under:
:
:
appl1
anim.wg
3.18.
With manual start selected, the
averaging routine begins when
you press a key (provided the start
conditions are met)� With automatic
start selected, averaging begins when
you place the first load on the platform
(provided the start conditions are met)�
– Enter the number of subweighing
operations using the keypad
– Info mode
– Toggle the display from "result of last
measurement" to "current weight" by
pressing the
w
key
– Autmatic printout configured in Setup
under:
:
Appl1
:
anim.wg
3.20.
– Automatic taring of container weight�
Configured in Setup under:
appl1
:
anim.wg
:
3.7.
– Automatic start of averaging when
the scale is turned on and a sample
placed on the platform (provided start
conditions are met)� Configured in
Setup under:
Appl
:
a.start
:
3.8.
– Exit application, delete parameters:
The number of measurements remain in
memory until it is deleted, overwritten
or the application is changed using the
c key�
The number of measurements also
remains saved after the scale is turned
off�
You can assign different functions to
the c key for deleting applications�
When you clear applications, you can
delete either the data stored for all
applications or just selected data stored
for the active application� Configured in
Setup under:
54
Signum 3 Ex
V
:
:
Appl
sel.cf
3.24.
(Sel� CF function key c in
applications)
– Tare function:
If you store a tare (weight value) by
pressing the ) key, you can later
enter a value manually� The tare value
you enter is added to the stored tare
value�
Setting: menu code 3�25�1 (factory
setting)
A tare value entered manually
overwrites a stored tare value (weight
value)� If you enter a tare value
manually, a tare value (weight value)
stored later overwrites the manually
entered value� Setting: menu code
3�25�2
Configured in Setup under: Signum
Appl: tare. f: 3.25.
– Restore factory default settings�
Configured in Setup under:
Appl: DEF.App: 9.1.
A number of subweighing operations
are required to form the basis for
calculation of an average weight�
You can enter the desired number
of subweighing operations using the
keypad�
The number you enter is active until
it is overwritten by another number�
It also remains in memory when you
switch to a different application
program, or turn off the scale�
There are three ways to start the
averaging routine:
– Manual start with preset number of
sub-weighing operations:
Place the sample on the platform and
press the O key
®
2
– Manual start with user-defined number
of subweighing operations:
Place the sample on the platform
and enter the number of weighing
operations using the keypad� Press the
r key to save the number entered
and begin weighing
– Automatic start with preset number of
subweighing operations:
Measurement begins when you place
the first sample on the platform,
provided the start conditions are met�

Hide quick links: