Option 002 Power Cal:
Figure 15-2
NOTE
NOTE
Adjustments
1. Connect the equipment as shown:
Option 002 Power Call Connection
The DUT is connected to the Power Meter via the MSIB for the HP 70100A
but via HP-IB if the HP 437A/B is used.
The sensor is connected to the RF Output if the HP 70428A Option 002
Option 002 with only the F-F adapter, HP part number 5061-5311 or
equivalent.
2. Begin the power calibration of YTF2, by pressing the following
softkeys:
Service
YTF Cal
YTF2 Power
3. The process takes approximately 15 minutes to complete. Once
complete, the unit will have accurately determined the detector
calibration to set output power.
15-4 HP 70427A/HP 70428A User's Guide