Procedure
Performance Tests
Test 5 - Mixer Linearity (IF)
Step 1 - Set up the measuring receiver
Frequency: 600 MHz manual tune
AM mode
Average detector
20 kHz low-pass filter
2.1
SPCL
6.1
SPCL
Step 2 - Initialize the HP 70427A
Press the
INSTR PRESET
Step 3 - Set the HP 70427A input frequency to 18 GHz
1. Press the
softkey.
Output
2. Press the
softkey.
FREQ
3. Enter 18 with the numeric keypad.
4. Complete the entry by pressing the
Step 4 - Set the HP 70427A input attenuator to 10 dB
1. Press the
Amplitudes
2. Press the
more 1 of 2
3. Press the
INPUT ATTEN
4. Enter the 10 with the numeric keypad.
5. Complete the entry by pressing the
Step 5 - Set up the sweep oscillator
Frequency: 18 GHz
Power: –2.5 dBm
Modulation: External AM
Step 6 - Set up the pulse/function generator
Frequency: 10 kHz
Waveshape: Square
Level: Adjust for about 33% ± 0.5% AM measured on the
measuring receiver.
14-36 HP 70427A/HP 70428A User's Guide
(selects 40% AM range)
(selects fast AM ALC)
key.
softkey.
GHz
softkey.
softkey.
softkey.
softkey.
dB