Overview
Procedure
Figure 2-48
NOTE
Getting Started
To Output a Signal with the Tunable Source
To Output a Signal with the Tunable Source
The procedure describes how to up-convert the output of an HP 8662A or
HP 8663A signal generator with the HP 70428A Option 002.
Step 1 - Connect the Signal Generator.
Signal Generator Connection
1. Connect an HP-IB cable between the signal generator and the MMS
mainframe that the HP 70428A Option 002 is installed in.
2. Connect an RF cable between the signal generator's RF OUTPUT
connector and the HP 70428A Option 002's HP 8662A connector.
The HP 70428A Option 002 acts as a controller. There must be no other
controller connected to the bus.
2-44 HP 70427A/HP 70428A User's Guide