Rotary Pump Exhaust Outlet; Inlet System Exhaust - Thermo Scientific Prima PRO User Manual

Mass spectrometers
Table of Contents

Advertisement

Temperatures
Rotary Pump
Exhaust Outlet
Inlet System
Thermo Fisher Scientific
High
Warning! The following assemblies may attain a surface
temperature as listed:
Warning! Ensure that the rotary pump is unobstructed. Failure to
do so will cause dangerous internal pressures in the pump. This will
result in blown seals and bursting of the oil box and could result in
physical injury.
Exhaust
Warning! A leak-tight exhaust must be fitted to the inlet system
when sampling gas streams which are:
Rapid Multi-Stream Sampler (RMS) and Single Point Inlet
System: 120°C.
Single Point Inlet System (High Temperature Version):
230°C.
RMS Sample Probe: 120°C.
Inlet Probe Assembly: 120°C.
Rotary pump operating temperature: 75°C.
Explosive
Toxic
Asphyxiant
Prima PRO & Sentinel PRO Mass Spectrometers User Guide
Safety Information
System Safety
2-5