Startup & Shutdown
Procedure
Mass Alignment
7-8
Prima PRO & Sentinel PRO Mass Spectrometers User Guide
Figure 7–8. Sensitivity tab
Detectors
The Detectors tab
systems with both Faraday and Secondary Emission Multiplier
(SEM) detectors.
As both detectors are off-axis, a deflection voltage is applied to
divert the ion beam into the selected detector. Deflector voltage
settings for both detectors are optimized on this tab.
The SEM voltage can be adjusted (i.e. calibrated) to set a specified
gain (typically 103 or 104).
Figure 7–9. Detectors tab
(Figure
7–8) is only available for dual detector
Thermo Fisher Scientific