Detectors; Mass Alignment - Thermo Scientific Prima PRO User Manual

Mass spectrometers
Table of Contents

Advertisement

Startup & Shutdown
Procedure

Mass Alignment

7-8
Prima PRO & Sentinel PRO Mass Spectrometers User Guide
Figure 7–8. Sensitivity tab

Detectors

The Detectors tab
systems with both Faraday and Secondary Emission Multiplier
(SEM) detectors.
As both detectors are off-axis, a deflection voltage is applied to
divert the ion beam into the selected detector. Deflector voltage
settings for both detectors are optimized on this tab.
The SEM voltage can be adjusted (i.e. calibrated) to set a specified
gain (typically 103 or 104).
Figure 7–9. Detectors tab
(Figure
7–8) is only available for dual detector