Prima PRO & Sentinel PRO Mass Spectrometers User Guide
The Heaters tab
the source temperature, inlet probe heater power, RMS body

Heaters

temperature, and RMS sample tube temperature.
Figure 7–11. Heaters tab
Open the Instrument Status window (see
the readings displayed under each tab are listed below.
Status
Analyser tab
Source tab
(Figure
7–11) allows for adjustment and setting of
Cabinet temperature: Within the range 10°C to 35°C.
Electronics temperature: Within the range 10°C to 45°C.
System pressure: Typically 5 x 10
dependent. Refer to installation records for setting of
optimum pressure.
Ion energy: Approximately 1000 V (for 150 amu mass range)
or 800 V (for 200 amu mass range).
Electron energy: Typically set to 70 V.
Filament current: Depends on filament type:
2.4-2.9 A for a Thoria coated iridium filament
1.5-2.5 A for a W/ThO coiled filament
3.0- 3.8 A for a Re filament
Note If filament is close to zero, check that Trap Current is
set to at least 20 µA and EE is at least 50 V.