Performing the DC Subsystem Functional Verification Test

Complete these steps to run the self- test and verify the operation the DC
subsystem.
For Agilent 4156C subsystems:
To perform the Agilent 4156C precision semiconductor parameter analyzer self-test

If you encounter a problem

Installation and User's Guide
Step
1 Disconnect all cables from the
measurement terminals on the
4156C rear panel.
2 Connect the 4156C to line power.
3 Switch on the instrument line
power switch.
4 Wait 1 hour before continuing to
step
5.
5 Press System > [CALIB/DIAG].
6 Press [DIAG SELFTST ALL].
Refer to the Agilent 4156C precision semiconductor parameter analyzer
manual titled, "If You Have a Problem" for an explanation of the error
codes. See the chapter titled, "If Errors Occur When You Perform
Self- Calibration or Diagnostics."
Refer to the troubleshooting information in the service manual for the
Agilent 4156C precision semiconductor parameter analyzer.
DC Subsystem Functional Verification Test
Notes
The analyzer displays the SYSTEM:
SELF-CALIBRATION/DIAGNOSTICS
screen.
This begins the calibration and
self-test process.
PASS, FAIL, or DONE appear in the
STATUS column.
If a failure occurs, an error code is
displayed in the ERROR column. See
"If you encounter a
If no errors occur, DIAG SELF-TEST
ALL: PASS appears in the lower
left-hand corner of the display.
B
problem."
121