To Troubleshoot The System - Agilent Technologies 85225F Installation And User Manual

Performance modeling system
Table of Contents

Advertisement

4
Servicing the System

To troubleshoot the system

Most system problems are caused by faulty cabling or switch
configurations.
Refer to
following:
If the cabling and switch configurations are verified correct, do the
following:
If a problem with one of the system components is found, refer to the
troubleshooting and repair information in the individual instrument's
product documentation.
For more information, see
104
Figure 20
on page 53 and
Check system connections and settings:
1
system connections to the DUT
2
system interconnections
3
GPIB cabling
4
GPIB address settings
Complete the Agilent E8364B PNA Series vector network analyzer
operator's check in
Verification Test"
on page 125.
Complete the Agilent 4156C precision semiconductor parameter
analyzer self- test in
Verification Test"
on page 121.
Complete the Agilent 4284A self- test by cycling the instrument's line
power. If errors occur, refer to Appendix B in the precision LCR meter
operation manual. The operation manual is included in with the
Agilent 85225F performance modeling system.
If you suspect trouble with the E2050B, see "Chapter 4
Troubleshooting" in the E5810A LAN/GPIB gateway installation and
configuration guide.
Complete the Agilent E5250A self- test and leak test found in Chapter 3
of the low leakage switch mainframe user's guide. The user's guide is
included with the Agilent 85225F performance modeling system.
Figure 19
"Performing the RF Subsystem Functional
"Performing the DC Subsystem Functional
"To receive additional assistance"
on page 51 and do the
on page 107.
Installation and User's Guide