Installing the System

Agilent 4156C Systems with Agilent 41501B Expander Box

To connect the bias networks
Step
1 For Agilent 4156C with
41501B expander box
systems, refer to the
following figure.
2 Connect the triaxial
cables from the 4156C to
the bias networks.
3 Connect and route the
triaxial cable from 4156C
HRSMU3 FORCE
connector to the port 2
bias network.
82
Action
a Connect the triaxial cables from the 4156C
HRSMU1 FORCE and SENSE connectors to
the DC FORCE and DC SENSE connectors
on the port 1 bias network.
b Connect the triaxial cables from the 41501B
HPSMU FORCE and SENSE connectors to
the DC FORCE and DC SENSE connectors
on the port 2 bias network.
a From the rear of the system, connect one
end of the triaxial cable to the GNDU
connector on the 41501B rear panel.
b From the rear of the system, route the cable
through the upper feedthrough panel.
c Connect the triaxial cable to the port 2 bias
network GNDU connector.
Notes
From the rear of the system, route the
cable through the feedthrough panel.
The port 1 bias network is the 11612V
K11.
The port 2 bias network is the 11612V
K21.
Refer to
Figure
20, "Rear Panel Wiring
Diagram," on page 53.
The triaxial cable model number is
16494A Option 002.
Leave the GNDU connector on the port
1 bias network open.
Refer to
Figure
3, "Front Panel
Connections with Agilent 4156C," on
page 25.
Installation and User's Guide