Performance Characteristics And Specifications; Table 15. Supplemental System Characteristics; Interference Standards; Performance Modeling System Performance Specifications - Agilent Technologies 85225F Installation And User Manual

Performance modeling system
Table of Contents

Advertisement

Performance Characteristics and Specifications

Supplemental characteristics are not specifications, but are provided in
Table 15
Table 15 Supplemental System Characteristics
Characteristic
Line voltage
Line frequency
Circuit breaker amperage rating
Rack weight capacity
Rack external dimensions
Rack footprint (top view)

Interference Standards

The IEC/EN 61326- 1 and CISPR Publication 11 standards define the RFI
and EMI susceptibility of the performance modeling system.

Performance Modeling System Performance Specifications

The Agilent 85225F performance modeling system adheres to the
performance specifications of an Agilent E8364B PNA Series vector
network analyzer. Refer to
Specification Summary," starting on page 141.
Installation and User's Guide
Introducing the Agilent 85225F Performance Modeling System
for your convenience.
Value
115 volts nominal (90 volts to 132 volts) or 220 volts nominal
(210 volts to 250 volts)
48 Hz to 66 Hz
6 amps (115 volts), 3.5 amps (220 volts)
818 kilograms (1800 pounds) maximum loaded
1620 mm high 600 mm wide 905 mm deep