Agilent Technologies 85225F Installation And User Manual page 100

Performance modeling system
Table of Contents

Advertisement

3
Verifying System Functionality
To perform the system functional verification test (continued)
Step
15 For systems with the
Agilent 4284A precision
LCR meter, open the
junction capacitance
model (juncap.mdl) in
IC-CAP.
16 Select the DUT/Setup.
17 Set the LCR meter
instrument options.
18 Calibrate the LCR meter.
19 Measure the open circuit
capacitance.
20 Observe the results
* To familiarize yourself with the IC-CAP software, refer to the first three chapters of the Agilent IC-CAP 2004 User's Guide, model number
85190D.
† To save this hardware configuration: on the IC-CAP main menu bar, choose File > Save As and enter a filename, for example
config1.hwd (the file suffix must be .hwd).
‡ To save this hardware configuration: on the IC-CAP main menu bar, choose File > Save As and enter a filename, for example
config1.hwd (the file suffix must be .hwd).
**S12 is identical to S21 because the attenuator is assumed to be symmetrical.
This completes the functional verification procedure.
100
Action
a From the IC-CAP/Main window menu bar,
choose File > Examples... .
b In the Directories list of the File Open dialog
box, double-click on the directory
.../examples/model_files.
c In the Directories list of the File Open dialog
box, double-click on the directory
.../model_files/diode.
d In the Files list of the File Open dialog,
double-click on juncap.mdl.
a Select the DUTs-Setups tab.
b Click the area > cv setup.
a Select the Instrument Options tab.
b Highlight the Cable Length variable and
enter 2.
a On the 16048D test leads, connect one tee
between the H
and H
pot
b Connect the other tee between the L
L
connectors. Select the
cur
Measure/Simulate tab.
c Select Calibrate.
d Follow the IC-CAP prompts.
a Select Measure.
a After the calibration is complete, observe
the plot.
b Rescale the measured result (displayed in
red) cap.m.
Notes
This opens the File Open dialog box.
This opens a list of model files.
(Scroll the list, if needed.) This displays
a list of modeling files in the Files list of
the File Open dialog.
This opens the juncap model window.
The juncap model window contains tab
folders used to interact with the model
data.
This opens the capacitance vs. voltage
setup.
This sets the cable length in the LCR
meter instrument options.
This performs a calibration on the LCR
connectors.
meter.
cur
and
The BNC tee part number is 1250-2405.
pot
This measures the open circuit
capacitance.
The measured result should be less
than ±10E-15 farads.
Installation and User's Guide

Advertisement

Table of Contents
loading

Table of Contents

Save Article as PDF