Agilent Technologies 85225F Installation And User Manual page 96

Performance modeling system
Table of Contents

Advertisement

3
Verifying System Functionality
To perform the system functional verification test (continued)
Step
3 Start the IC-CAP
*
software.
4 Add the system interface
and components to the
IC-CAP Hardware Setup.
5 For 4156C systems,
change the Agilent 4156C
precision semiconductor
parameter analyzer
HRSMU names in IC-CAP.
6 For E5260A/70B systems,
change the Agilent 4156C
precision semiconductor
parameter analyzer SMU
names in IC-CAP.
96
Action
PC version:
a Select Start > Programs > IC-CAP 2004 >
IC-CAP.
UNIX version:
a Open a UNIX terminal window.
b At the prompt, type iccap.
c Press Enter.
a From the IC-CAP/Main window menu bar,
choose Tools > Hardware Setup... .
b If you have not added an interface, click
Add Interface, enter the system interface
(hpib or other), and click OK.
c In the Instrument List, click Rebuild.
d Verify that all system components appear
in the Instrument List.
a In the Instrument List, select HP4156 (hpib,
19) and click Configure... .
b In the Configuration of HP4156 window
Unit Table group box, highlight the
characters in the HRSMU1 entry box and
type VG.
c In the Unit Table group box, highlight the
characters in the HPSMU entry box and
type VD.
d Click OK.
e Close the Hardware Setup window.
a In the Instrument List, select Agilent E5270
(hpib, 19) and click Configure... .
b In the Configuration of E5270 window Unit
Table group box, highlight the characters in
the MPSMU<slot number> entry box and
type VG.
c In the Unit Table group box, highlight the
characters in the HPSMU<slot number>
entry box and type VD.
d Click OK.
e Close the Hardware Setup window.
Notes
This starts IC-CAP and opens the
IC-CAP/Status and IC-CAP/Main
windows.
This opens the IC-CAP/Hardware
Setup window.
This polls the GPIB and adds all
connected and activated system
instruments to the Instrument List.
Disregard error messages on the
system instrument displays. The errors
are a by-product of the GPIB polling
process.
This opens the Configuration of
HP4156 window where the Unit Table
dialog box is used to change the names
of the HRSMUs.
Actions b, c, and d change the name of
HRSMU1 to VG and HPSMU to VD.
Renaming the SMUs is necessary for
proper execution of the example model
file.
This opens the Configuration of E5270
window where the Unit Table dialog
box is used to change the names of the
SMUs.
Actions b, c, and d change the name of
MPSMU<slot number> to VG and
HPSMU<slot number> to VD.
Renaming the SMUs is necessary for
proper execution of the example model
file.
Installation and User's Guide