Replacing the Erase Lamp Inverters

Laser Warning
To avoid possible eye damage, before operating the unit without the cover, disconnect the laser (see
the Laser" on Page
35). For additional laser precautions see
personnel may remove the cover.
[1] Remove the Main Cover (see

Inverter Cover Screws

[2] Remove the 4 screws securing the Inverter Cover.

Erase Lamp Connectors

[3] Disconnect the Erase Lamps Connectors from the Inverters.
[4] Disconnect the cables from the Erase Lamp Board.
Catalog No. 6H1344 – 10APR2006
"Laser Safety Instructions" on Page
"Removing the Main Cover" on Page
Replacing Components
11. Only authorized
33).
"Disconnecting
57