SERVICE GUIDE
• Four 3.0 mm screws
• Sensor Cable
• Motion Board Cable
[2] Close the Access Panel and return the Scanner on to its base.
[3] Reconnect the Sensor Cable to the Sensor Board and the Motion Cable to the Motion Board and fasten the
cable ties.
[4] Perform the test procedures in the section below before replacing the Scanner Cover.
[5] Replace:
• Scanner Cover
• Service Panel
Test Procedures
Laser Warning
When a service operation is taking place with the cover removed, disconnect the Laser according to the procedure
in
"Disconnecting the Laser" on Page
safety glasses at all times.
[1] Confirm the Loader Pin height is correct:
(a) Turn on the Scanner.
(b) Log in into Technician mode.
(c) Choose "Setup" and enter the "Diagnostics" tab.

Diagnostics Screen Loader Control

(d) Click on the [<<] sign to move the loader back to the end of its travel and the "loader back" indicator lights.
46
35. If the Laser must be activated during the service procedure, wear protective
10APR2006 – Catalog No. 6H1344