Inverter Assembly

The Inverter Assembly converts the 15 VDC input to a high voltage output to power the Erase Lamps.

Inverter Assembly

Inverter Assembly Connections

Inverter Assembly Voltages
Connector
Catalog No. 6H1344 – 10APR2006
1
J1 ON Lamp Sensor Board
1
2
3
4
2
3
1
2
4
AC - High voltage to Erase Lamps
Destination
Red
White
Lamp on 0 V
Lamp off 2.5 - 5 V
Green
Lamp on 0 V
Lamp off 2.5 - 5 V
Black
GND
J202 on motion board
P2 on 15 V power supply
Red
15 V
White
GND
Component Description
5 V