Replacing the Left Limit Sensor

Laser Warning
To avoid possible eye damage, before operating the unit without the cover, disconnect the laser (see
the Laser" on Page
35). For additional laser precautions see
personnel may remove the cover.
[1] Remove the Main Cover (see
Left Limit Sensor Connector
[2] Disconnect the Sensor Cable from the Connector Panel.
[3] Mark the location of the sensor.

Left Limit Sensor Screws

[4] Remove the 2 screws attaching the sensor to the sensor bracket, and remove the sensor.
[5] Reverse the above steps, to install the Left Limit Sensor, using the marked location of the previous sensor.