Replacing the Slide

Laser Warning
To avoid possible eye damage, before operating the unit without the cover, disconnect the laser (see
the Laser" on Page
35). For additional laser precautions see
personnel may remove the cover.
[1] Using the user interface Diagnostic screen, move the Optical Head to the middle of the drum.
[2] Remove the main cover (see

Connector to USB Board

[3] Disconnect the flex cable from the USB Board.

Laser Board Cable and Connectors

[4] Disconnect the Laser Board flex cable from the Motion Board.
[5] Release the 2 flex cables, until they can be taken out freely along with the slide.
Catalog No. 6H1344 – 10APR2006
"Removing the Main Cover" on Page
"Laser Safety Instructions" on Page
33).
Replacing Components
"Disconnecting
11. Only authorized