31
: : Walking (repeat for readjustment)
Since the switching threshold buffer exhibits settlement (especially during the first 10 minutes), it
is essential to check again while walking on a level surface whether short, quick steps are still
possible or whether the settings have to be readjusted.
: Check the swing phase flexion resistance again and, if necessary, adjust it in small incre
►
6.1
ments so the lower leg of the prosthesis does not swing through too far in dorsal position and
reaches full extension in time for the next heel strike.
: During the exercise, also check whether the swing phase extension resistance setting is
►
6.2
still appropriate and adjust it if necessary.
: After changing the swing phase extension resistance, check the switching threshold set
►
6.3
ting and adjust it if necessary.
: After changing the static alignment during dynamic fitting (e.g. increasing plantar flexion),
►
6.3
check the switching threshold setting and adjust it if necessary.
31
6.1
6.2
6.3
39