Installing the tube adapter
CAUTION
Improper assembly of the screw connections
Risk of injury due to breakage or loosening of the screw connections
Clean the threads before every installation.
►
Apply the specified torque values.
►
Follow the instructions for thread lock.
►
CAUTION
Incorrect processing of the tube
Fall due to damage to the tube
Do not clamp the tube in a vice.
►
To shorten the tube, use only a tube cutter or a cutting device.
►
Deburr the inside and outside of the cut edge with the tube deburrer.
►
CAUTION
Incorrect assembly of the tube
Risk of injury due to breakage of load-bearing components
Clean the contact surfaces of the tube and the tube adapter using a degreasing cleaning
►
agent.
Note the values identified as permissible and impermissible in the illustration for positioning
►
the tube in the tube adapter.
17