Installing the tube adapter
CAUTION
Improper assembly of the screw connections
Risk of injury due to breakage or loosening of the screw connections
Clean the threads before every installation.
Apply the specified torque values.
Follow the instructions for thread lock.
CAUTION
Incorrect processing of the tube
Fall due to damage to the tube
Do not clamp the tube in a vice.
To shorten the tube, use only a tube cutter or a cutting device.
Deburr the inside and outside of the cut edge with the tube deburrer.
CAUTION
Incorrect assembly of the tube
Risk of injury due to breakage of load-bearing components
Clean the contact surfaces of the tube and the tube adapter using a degreasing cleaning
agent.
Note the values identified as permissible and impermissible in the illustration for positioning
the tube in the tube adapter.