Syntax
:SOURce<n>:PATTern<m>:DEEMphasis:CURSor<wsp><0|POST|1|PRE|DEF>
Description
Set the pattern de-emphasis cursor.
Parameters
PRE: Set to pre-cursor
POST: Set to post-cursor
DEF: Restores the default value
Response
No response
Example
:SOURce3:PATTern2:DEEMphasis:CURSor POST
Command
:SOURce<n>:PATTern<m>:LOGic?
Syntax
:SOURce<n>:PATTern<m>:LOGic?[<wsp><DEF|SET|ALL|LIST|INFO>]
Description
Query the data pattern level logic
Parameters
DEF: Query the default power-on value
SET: Query the current SET value
LIST: Query the list of discrete options for configurable values
ALL: Query the values specified by DEF, SET, and LIST
INFO: Provides a description of what the return values represent
By default, the command Query the current SET value.
Response
An integer indicating the pattern level logic.
Example
:SOURce5:PATTern1:LOGic? ->1
:SOURce5:PATTern1:LOGic? INFO ->
0:LOW,NEGATIVE
1:HIGH,POSITIVE
Command
:SOURce<n>:PATTern<m>:LOGic
Syntax
:SOURce<n>:PATTern<m>:LOGic<wsp><0|NEGative|LOW|1|POSitive|HIGH|DEF>
Description
Set the data pattern digital logic level.
Parameters
value: The desired new set point
DEF: Restores the default value
Response
No response
Example
:SOURce3:DATA2:LOGIc POSITIVE
:SOURce<n>:PATTern<m>:TYPE?
Command
Syntax
:SOURce<n>:PATTern<m>:TYPE?<wsp><MIN|MAX|DEF|SET|LIST|ALL|INFO|LINKED>
Description
Query the pattern type.
Parameters
DEF: Query the default power-on value
SET: Query the current SET value
LIST: Query the list of discrete options for configurable values
ALL: Query the values specified by DEF, SET, and LIST
INFO: Provides a description of what the return values represent
LINKED: Query the linked channels
By default, the command Query the current SET value.
Response
A single integer value, or a comma separated integer array. When ALL is requested, the
format is "<DEF>,<SET>,<LIST>".
:SOURce3:PATTern2:TYPE? -> 6
Example
Command
:SOURce<n>:PATTern<m>:TYPE
:SOURce<n>:PATTern<m>:TYPE<wsp><value|MIN|MAX|DEF>
Syntax
Description
Set the pattern type.
Parameters
value: The desired new set point
MIN: Set to the minimum value
MAX: Set to the maximum value
DEF: Restores the default value
Response
No response
Example
:SOURce3:PATTern2:TYPE PRBS9