Quantifi Photonics BERT 1005 Series User Manual page 53

Bit error rate tester
Table of Contents

Advertisement

8.5.1.1
Set the cursor and de-emphasis
To set the pre-cursor or post-cursor de-emphasis, follow the steps bellow:
1.
Select an appropriate value from the DE-EMPHASIS drop-down list.
2.
Select PRE or POST from the CURSOR drop-down list.
8.5.1.2
Example of setting the pre-cursor and post-cursor de-emphasis
Consider the following pre-cursor and post-cursor example:
Pre-De-emphasis is set to a lower number while it is x for the post-De-emphasis
The Peak Swing is set by the AMPLITUDE field. The de-emphasis control leaves the Peak Swing
unchanged, but it reduces the De-emphasized Swing, to provide the desired level of de-emphasis.
The De-emphasis swing is calculated as �� = 20 × log (
< m > represents the Peak Swing, and < n > represents the De-emphasized Swing.
For example, if the value of the AMPLITUDE field (Peak Swing) is set to 800 mV, and the value of De-
emphasized Swing is 505 mV, it can be calculated that the value of the DE-EMPHASIS field must be set
to 4 dB.
The amplitude of the 0 and 1 levels can be calculated by dividing the AMPLITUDE (Peak Swing) by 10 to
the power of the DE-EMPHASIS divided by 20. The difference between the 0 and 1 levels for the above
800mV
example would be: �� = (
Quantifi Photonics Ltd.
) = 505����
4����
10
20
Version
Bit Error Rate Tester | BERT 1005
��
), where < d > represents the de-emphasis,
��
2.04
53