Quantifi Photonics BERT 1005 Series User Manual page 73

Bit error rate tester
Table of Contents

Advertisement

Syntax
:SENSe<n>:MEASure<m>:EDIagram:HORIzontal<wsp><value|MIN|MAX|DEF>
Description
Set the horizontal (x) resolution step size of the eye measurement.
Parameters
value: The desired new set point
MIN: Set to the minimum value
MAX: Set to the maximum value
DEF: Restores the default value
Response
No response
Example
:SENSe3:MEASure2:EDIagram:HORIzontalstep 4
Command
:SENSe<n>:MEASure<m>:EDIagram:PatternLENgth?
Syntax
:SENSe<n>:MEASure<m>:EDIagram:PatternLENgth?[<wsp><MIN|MAX|DEF|SET|
LIST|ALL|LINKED>]
Description
Query the eye measurement pattern length (2^n bits) resolution.
Parameters
MIN: Query the minimum configurable value
MAX: Query the maximum configurable value
DEF: Query the default power-on value
SET: Query the current SET value
LIST: Query the list of discrete options for configurable values (if available)
ALL: Query the values specified by MIN, MAX, DEF, SET, and LIST
LINKED: Query the linked channels
By default, the command Query the current SET value.
Response
A single value, or a comma separated array. When ALL is requested, the format is
"<MIN>,<MAX>,<DEF>,<SET>,<LIST>".
Example
:SENSe3:MEASure2:EDIagram:PatternLENgth? MAX -> 16
Command
:SENSe<n>:MEASure<m>:EDIagram:PatternLENgth
Syntax
:SENSe<n>:MEASure<m>:EDIagram:PatternLENgth<wsp><value|MIN|MAX|DEF>
Description
Set the eye measurement pattern length (2^n bits) resolution.
Parameters
value: The desired new set point
MIN: Set to the minimum value
MAX: Set to the maximum value
DEF: Restores the default value
Response
No response
Example
:SENSe3:MEASure2:EDIagram:PatternLENgth 16
Command
:SENSe<n>:MEASure<m>:EDIagram:STARt
Syntax
:SENSe<n>:MEASure<m>:EDIagram:STARt
Description
Starts the eye scan measurement. This command must always be used in conjunction
with the :CALCulate<n>:DATA<m>:EDIagram:STATus? query to ensure that the ED
measurements are not running.
Parameters
No parameters
Response
No response
:SENSe3:MEASure2:EDIagram:STARt
Example
Command
:SENSe<n>:MEASure<m>:EDIagram:STOP
Syntax
:SENSe<n>:MEASure<m>:EDIagram:STOP
Description
Stops a running eye scan measurement.
Parameters
No parameters
Response
No response
Example
:SENSe<n>:MEASure2:EDIagram:STOP
Command
:SENSe<n>:MEASure<m>:EDIagram:VERTical?
Quantifi Photonics Ltd.
Version
2.04
Bit Error Rate Tester | BERT 1005
73

Advertisement

Table of Contents
loading

Related Products for Quantifi Photonics BERT 1005 Series

Table of Contents