Specific Command Summary - Quantifi Photonics BERT 1005 Series User Manual

Bit error rate tester
Table of Contents

Advertisement

Command
*OPC?
Syntax
*OPC?
Description
Query the Operation Complete Status
Parameters
No parameters
is returned if all the modules installed in the chassis are ready to execute commands
1
Response
is returned if any module in the chassis has a command to execute in the input queue
0
Example
*OPC? -> 1
Command
*OPT?
Syntax
*OPT?
Description
Query the modules managed by the CohesionSCPI service
Parameters
No parameters
Response
Response will be a comma separated string of the installed modules in the chassis
*OPT? -> ,LaserPXIe-1002-2-FA,SwitchPXIe-1003-1-FC,,VOAPXIe-1001-2-
Example
FA,,,,O2EPXIe-1001-1-FC,,,,,,,,,

Specific command summary

Slot commands
:SLOT<n>
:OPC?
:TeST?
:OPTions?
:IDN?
:GRouP?
:PPG?
:ED?
:TEMPerature?
Configuration commands
:CALCulate<n>
:DATA<m>
:EALarm?
:ELAPsed?
:EDIagram?
:STATus?
:BATHtub?
:STATus?
:ELAPsed?
:Q?
:BERFloor?
:CENTer?
:JITTer?
:OUTPut<n>
:CLOCk
:VCOLock?
:DIVIde?
:DIVIde
:FREQuency
:ARBitrary?
:ARBitrary
:STanDard?
:STanDard
Quantifi Photonics Ltd.
Description
- Query the status of the Operation Complete bit
- Perform a self-test of the module
- Query the modules installed on the slot
- Query the Identifier for the slot; returns the manufacturer, part
number, serial number, hardware and firmware versions
- Query the PPG channels within the specified group
- Query the ED channels within the specified group
- Query the temperature of all groups of channels
Description
- Query the ED error measurement
- Query the running time for the ED measurement
- Query the eye scan measurement data
- Query the eye scan measurement status
- Query the BER values from the bathtub measurement
- Query the status of the bathtub measurement
- Query the elapsed time of the bathtub measurement and the time to
next point
- Query the Q values from the bathtub measurement
- Query the BER floor value from the bathtub measurement
- Query the center of the eye from the bathtub measurement
- Query the Jitter values from the bathtub measurement
- Query the stability of the clock synthesizer
- Query the clock synthesizer divide ratio
- Set the clock synthesizer divide ratio
- Query the current synthesizer frequency arbitrary settings
- Set the current synthesizer frequency to an arbitrary value within the
valid operating range
- Query the current synthesizer frequency standard settings
- Set the current synthesizer frequency to a standard value
Version
2.04