Quantifi Photonics BERT 1005 Series User Manual page 65

Bit error rate tester
Table of Contents

Advertisement

Description
Set the synthesizer frequency to an arbitrary value within the valid operating range.
Parameters
value: The desired new value
MIN: Set to the minimum value
MAX: Set to the maximum value
DEF: Restores the default value
Response
No response
Example
:OUTP3:CLOC:FREQ:ARB 7.96
Command
:OUTPut<n>:CLOCk:FREQuency:STanDard?
Syntax
:OUTPut<n>:CLOCk:FREQuency:STanDard?[<wsp><MIN|MAX|DEF|SET|LIST|ALL|
STEP|UNIT>]
Description
Query the current synthesizer frequency standard settings.
Parameters
MIN: Query the minimum configurable value
MAX: Query the maximum configurable value
DEF: Query the default power-on value
SET: Query the current SET value
LIST: Query the list of discrete options for configurable values (if available)
ALL: Query all the values specified by MIN, MAX, DEF, SET, and LIST
STEP: Query the minimum resolution of the frequency value
UNIT: Query the default unit for the frequency value
By default, this command Query the current SET value.
Response
A single value or a comma separated string of values. For ALL, the format of the
response is "<MIN>,<MAX>,<DEF>,<SET>,<LIST>".
If the frequency is set to an arbitrary value, the SET value in the response is NAN.
Example
STANDARD MODE:
:OUTP3:CLOC:FREQ:STD? -> 14.5
ARBITRARY MODE:
:OUTP3:CLOC:FREQ:STD? -> NAN
Command
:OUTPut<n>:CLOCk:FREQuency:STanDard
Syntax
:OUTPut<n>:CLOCk:FREQuency:STanDard<wsp><value|MIN|MAX|DEF>
Description
Set the current synthesizer frequency to a standard value.
Parameters
value: The desired new value
MIN: Set to the minimum value
MAX: Set to the maximum value
DEF: Restores the default value
Response
No response
Example
:OUTP3:CLOC:FREQ:STD 14.5
Command
:OUTPut<n>:CLOCk:POWer?
Syntax
:OUTPut<n>:CLOCk:POWer?[<wsp><MIN|MAX|DEF|SET|LIST|ALL|STEP|UNIT>]
Description
Query the power of the RF output clock
Parameters
MIN: Query the minimum configurable value
MAX: Query the maximum configurable value
DEF: Query the default power-on value
SET: Query the current SET value
LIST: Query the list of discrete options for configurable values (if available)
ALL: Query all the values specified by MIN, MAX, DEF, SET, and LIST
STEP: Query the minimum resolution of the power value
UNIT: Query the default unit for the power value
By default, the command Query the current SET value.
Response
A single value or a comma separated string of values. For ALL, the format of the
response is "<MIN>,<MAX>,<DEF>,<SET>,<LIST>".
Example
:OUTPut3:CLOCk:POWer? -> 0
Command
:OUTPut<n>:CLOCk:POWer
Syntax
:OUTPut<n>:CLOCk:POWer<wsp><value|MIN|MAX|DEF>
Quantifi Photonics Ltd.
Version
2.04