Description
Set the power for the RF output clock.
Parameters
value: The desired new set point
MIN: Set to the minimum value
MAX: Set to the maximum value
DEF: Restores the default value
Response
No response
Example
:OUTPut<n>:CLOCk:POWer 0.0
Command
:OUTPut<n>:DATA<m>:AMPlitude?
Syntax
:OUTPut<n>:DATA<m>:AMPlitude?[<wsp><MIN|MAX|DEF|SET|ALL|LIST|STEP|
UNIT|LINKED>]
Description
Query the voltage amplitude of the output signal.
Parameters
MIN: Query the minimum configurable value
MAX: Query the maximum configurable value
DEF: Query the default power-on value
SET: Query the current SET value
LIST: Query the list of discrete options for configurable values (if available)
ALL: Query the values specified by MIN, MAX, DEF, SET, and LIST
STEP: Query the minimum resolution of the amplitude value
UNIT: Query the default unit for the amplitude value
LINKED: Query the linked channels.
By default, this command Query the current SET value.
Response
A single value, or a comma separated array. When ALL is requested, the format is
"<MIN>,<MAX>,<DEF>,<SET>,<LIST>"
:OUTPut3:DATA2:AMPlitude? ALL ->200,1100,200,200,NAN
Example
This command does not have a discrete LIST.
Command
:OUTPut<n>:DATA<m>:AMPlitude
Syntax
:OUTPut<n>:DATA<m>:AMPlitude<wsp><value|MIN|MAX|DEF>
Description
Set the amplitude of the output signal.
Parameters
value: The desired new set point
MIN: Set to the minimum value
MAX: Set to the maximum value
DEF: Restores the default value
Response
No response
Example
:OUTPut3:DATA2:AMPlitude 0.60
Command
:OUTPut<n>:DATA<m>:CPOint?
Syntax
:OUTPut<n>:DATA<m>:CPOint?[<wsp><MIN|MAX|DEF|SET|LIST|ALL>]
Description
Query the pattern cross point (%).
Parameters
MIN: Query the minimum configurable value
MAX: Query the maximum configurable value
DEF: Query the default power on value
SET: Query the current SET value
LIST: Query the list of discrete options for configurable values (if available)
ALL: Query the values specified by MIN, MAX, DEF, SET, and LIST.
By default, the command Query the current SET value.
Response
A single value, or a comma separated array. When ALL is requested, the format is
"<MIN>,<MAX>,<DEF>,<SET>,<LIST>".
Example
:OUTPut<n>:DATA<m>:CPOint? LIST -> 35,40,45,50,55
Command
:OUTPut<n>:DATA<m>:CPOint
Syntax
:OUTPut<n>:DATA<m>:CPOint<wsp><value|MIN|MAX|DEF>