Switching From A Mymode Back To Basic Mode - Otto Bock Genium X3 3B5-3 Instructions For Use Manual

Hide thumbs Also See for Genium X3 3B5-3:
Table of Contents

Advertisement

Available languages

Available languages

MyModes
Switching process
3x
à MyMode 1
4x à MyMode 2
ð
1) Position the prosthetic leg slightly to the rear (step position).
2) While maintaining constant contact with the floor, bounce on the forefoot with the leg extended a number of
times in one second depending on the desired MyMode (MyMode 1 = 3 times, MyMode 2 = 4 times).
3) Fully unload and keep the prosthetic leg still in this position (step position).
→ A beep and vibration signal will sound to confirm that the movement pattern has been recognised.
INFORMATION: If this beep and vibration signal does not sound, the requirements were not
observed when bouncing the foot or mute mode (silent mode) is activated. For more information
about mute mode, see the section "Mute mode" (see page 95).
4) After the beep and vibration signal sounds, keep the prosthetic leg extended and still for 1 second.
→ A confirmation signal will sound to indicate that the prosthesis has successfully switched to the corresponding
MyMode (2 times = MyMode 1, 3 times = MyMode 2).
INFORMATION: If this confirmation signal does not sound, the leg with the prosthesis was not cor­
rectly kept still or mute mode (silent mode) is activated. Repeat the process for correct switching. For
more information about mute mode, see the section "Mute mode" (see page 95).

10.4 Switching from a MyMode back to basic mode

Information on switching
Regardless of the configuration of additional MyModes in the adjustment software, it is always possible to
switch back to basic mode (mode 1) with a motion pattern.
It is always possible to switch back to basic mode (mode 1) by connecting/disconnecting the battery charger.
Before the first step, always check whether the selected mode corresponds to the required motion type.
Requirements for successful switching using motion patterns
The following points must be observed to carry out switching successfully:
Position the prosthetic leg back slightly and bounce on the forefoot with the leg extended while maintaining
constant contact with the floor.
Weight must be placed on the forefoot while bouncing.
Do not take the weight off fully during unloading while bouncing.
Switching process
≥3x
ð
1) Position the prosthetic leg slightly to the rear (step position).
2) While maintaining constant contact with the floor, and with the leg extended, bounce on the forefoot three or
more times.
3) Fully unload and keep the prosthetic leg still in this position (step position).
→ A beep and vibration signal will sound to confirm that the movement pattern has been recognised.
INFORMATION: If this beep and vibration signal does not sound, the requirements were not
observed when bouncing the foot or mute mode (silent mode) is activated. For more information
about mute mode, see the section "Mute mode" (see page 95).
4) After the beep and vibration signal sounds, keep the prosthetic leg extended and still for 1 second.
98
1x
+
ð
1x
+
ð
1s
1x
ð ð
1s
1x
1x
ð ð
2x à MyMode 1
3x à MyMode 2
MyMode 1
+
ð
MyMode 2
1x
Basic mode
+
ð
þ
þ
Genium X3 3B5-3/3B5-3=ST
þ

Advertisement

Chapters

Table of Contents
loading

This manual is also suitable for:

Genium x3 3b5-3-st

Table of Contents