Display Of The Current Charge Level; Display Of Battery Charge Level Without Additional Devices; Display Of The Current Charge Level Using The Cockpit App; Display Of The Current Charge Level During The Charging Process - Otto Bock Genium 3B1-3 Instructions For Use Manual

Hide thumbs Also See for Genium 3B1-3:
Table of Contents

Advertisement

INFORMATION
Charging the product without tube adapter
When the inductive charger is disconnected from the knee joint without the tube adapter, five beep and vibration
signals are output. In order to avoid this, perform the following steps before disconnecting the charger:
► Hold the joint upright and flex the knee head 90°.
► Wait 5 seconds. The joint switches to energy saving mode.
→ The tube adapter can be disconnected without generating a warning signal.

6.3 Display of the current charge level

6.3.1 Display of battery charge level without additional devices

INFORMATION
The charge level cannot be displayed during the charging process, e.g. by turning the prosthesis over. The
product is in charging mode.
Beep signal
5x short
4x short
3x short
2x short
1x short
INFORMATION
If the Volume parameter is set to '0' in the Cockpit app (see Page 38) or if mute mode (silent mode) is activated,
there are no beep signals.

6.3.2 Display of the current charge level using the Cockpit app

Once the Cockpit app has been started, the current charge level is displayed in the bottom line of the screen:

6.3.3 Display of the current charge level during the charging process

During the charging process, the current battery charge level is indicated by the number of LEDs lit on the side of
the charger.
Genium 3B1-3/3B1-3=ST
1) Turn the prosthesis by 180° (the sole of the foot must face up).
2) Hold still for 2 seconds and wait for beep signals.
1.
38% – Charge level of battery for currently connected component
Battery charge level
more than 80%
60% to 80%
40% to 60%
20% to 40%
less than 20%
Charging the battery
25

Advertisement

Table of Contents
loading

Table of Contents