Technical Data - Otto Bock Genium X3 3B5-3 Instructions For Use Manual

Hide thumbs Also See for Genium X3 3B5-3:
Table of Contents

Advertisement

Available languages

Available languages

Technical data

(2) this device must accept any interference, including interference that may cause undesired operation of this
device.
L' utilisation de ce dispositif est autorisée seulement aux conditions suivantes:
(1) il ne doit pas produire d'interference et
(2) l' utilisateur du dispositif doit étre prêt à accepter toute interference radioélectrique reçu, même si celle-ci est
susceptible de compromettre le fonctionnement du dispositif.
Caution: Exposure to Radio Frequency Radiation.
The installer of this radio equipment must ensure that the antenna is located or pointed such that it does not emit
RF field in excess of Health Canada limits for the general population; consult Safety Code 6, obtainable from
Health Canada's website
http://www.hc-sc.gc.ca/rpb.
Responsible party:
Otto Bock Healthcare Canada Ltd.
5470 Harvester Road
L7L 5N5 Burlington, Ontario
Canada
Phone + 1-800-665-3327
Caution: Federal law (USA) restricts this device to sale by or on the order of a practitioner licensed by law of the
State in which he/she practices to use or order the use of the device.
16 Technical data
Ambient conditions
Transport in original packaging
Storage in the original packaging (≤3 months)
Long-term
storage
(>3 months)
Transport and storage between applications (without
packaging)
Operation
Time for warming to the operating temperature after
storage between applications, from -25 °C/-13 °F at an
ambient temperature of +20 °C/+68 °F
Time for cooling to the operating temperature after stor­
age between applications, from +70 °C/+158 °F at an
ambient temperature of +20 °C/+68 °F
Charging the battery
Product
Reference number
Mobility grade (MOBIS)
Maximum body weight including additional weight
Protection rating
Water resistance
Prox. system height to alignment reference point 3B5-3*
(pyramid)
Proximal system height up to alignment reference point
3B5-3=ST* (threaded connector)
102
in
the
original
packaging
-25 °C/-13 °F to +70 °C/+158 °F
-20 °C/-4 °F to +40 °C/+104 °F
Max. 93% relative humidity, non-condensing
-20 °C/-4 °F to +20 °C/+68 °F
Max. 93% relative humidity, non-condensing
-25 °C/-13 °F to +70 °C/158 °F
Max. 93% relative humidity, non-condensing
-10 °C/+14 °F to +60 °C/+140 °F
Max. 93% relative humidity, non-condensing
30 minutes
30 minutes
+10 °C/+50 °F to +45 °C/+113 °F
3B5-3*/3B5-3=ST*
3 and 4
150 kg
IP66/IP68
Maximum water depth: 3 m
Maximum time: 1 hour
Waterproof, corrosion-resistant, resistant to penetration
from jets of water
0 mm
26 mm
Genium X3 3B5-3/3B5-3=ST