ics are noted, taking additional steps may be necessary, for example changing the orientation or
moving the product 'X2 Bionic Prosthetic System 3B5-X2'
Table 6 – Recommended safe distances between portable and mobile HF telecommunica
tion equipment and the product 'X2 Bionic Prosthetic System 3B5-X2' – for non-life-sus
taining devices and systems
Directives and manufacturer's declaration – Recommended safe distances between
portable and mobile HF telecommunication equipment and the product 'X2 Bionic Pros
thetic System 3B5-X2'
The product 'X2 Bionic Prosthetic System 3B5-X2' is designed for operation in an electromag
netic environment where HF interference is monitored. The customer or user of the product 'X2
Bionic Prosthetic System 3B5-X2' can help prevent electromagnetic interference by complying
with the minimum distances between wearable and mobile HF communication equipment (trans
mitters) and the product 'X2 Bionic Prosthetic System 3B5-X2', as below according to the max
imum output of the communication equipment.
Nominal output of
transmitter [W]
0.01
0.1
1
10
100
For transmitters with no maximum nominal output specified in the table above, the recommen
ded safety distance d in metres (m) can be determined using the equation in the respective
column, where P stands for the maximum nominal output of the transmitter in watts (W) accord
ing to the transmitter manufacturer's information.
Note 1: The higher frequency range applies at 80 MHz and 800 MHz.
Note 2: These directives may not be applicable in all cases. The propagation of electromagnetic
factors is influenced by absorption and reflection by buildings, objects and people.
Safety distance [m] by transmission frequency
150 kHz to 80 MHz
N/A
N/A
N/A
N/A
N/A
80 MHz to 800 MHz
d = 0.3 √P
0.03
0.09
0.3
0.95
3
800 MHz to 2.5 GHz
d = 2.3 √P
0.23
0.73
2.3
7.3
23
107