Available languages

Available languages

Beep signal
5 x long
10 x long
30 x long
Error while charging the product
LED on
Status
power
LED on
supply
battery
charger
100
Vibration signal
3 x long
5 x long
5 x long repeated
every 60 seconds
10 x long
1 x long, 1 x short
repeated every
3 seconds
Continuous
Error
Country-specific plug adapter not
fully engaged on power supply
Event
Charge level under
25%
Charge level under
10%
Medium error (see
Page 92)
e.g. a sensor is not
operational
Battery charge level
5%
After the beep and
vibration signals, the
product switches to
empty battery mode
and then switches
off.
Severe error/indic­
ation of safety
mode activation
(see Page 92)
e.g. one or more
sensors are not oper­
ational
Total failure
Electronic control no
longer possible.
Safety mode active or
undetermined valve
state. Undetermined
product behaviour.
Resolution
Check whether the country-specif­
ic plug adapter is fully engaged on
the power supply.
Required action
Charge battery soon.
Remaining operating time
approx. 24 hours
Charge battery soon
Remaining operating time
approx. 6 hours
Walking possible with
restrictions. Please note
the change in flexion res­
istance.
The product must be
inspected immediately by
an O&P professional.
Charge the battery.
Attempt to reset this error
by connecting/discon­
necting the battery char­
ger.
If the error persists, use
of the product is prohib­
ited. The product must be
inspected immediately by
an O&P professional.
Attempt to reset this error
by connecting/discon­
necting the battery char­
ger.
If the error persists, use
of the product is prohib­
ited. The product must be
inspected immediately by
an O&P professional.