Otto Bock X2 Series Instructions For Use Manual page 101

Bionic prosthetic system
Hide thumbs Also See for X2 Series:
Table of Contents

Advertisement

Available languages

Available languages

LED on
Status
power
LED on
supply
battery
charger
Status
LED
The LED
ring is lit
in weak
purple
The LED
ring is lit
up yellow
The LED
ring is lit
up green
Error
Non-functional socket
Defective power supply
No connection between battery
charger and power supply
Defective battery charger
Charging status
indicator (5 LEDs)
No LED is lit
LED 2 and 4 are lit up Charger excess tem­
LED 1, 3 and 5 are lit
up
LED 3 is lit up
Resolution
Check socket with another electric
appliance.
The battery charger and power
supply must be inspected by an
authorised Ottobock Service
Centre.
Check whether the charging cable
plug is fully engaged on the battery
charger.
The battery charger and power
supply must be inspected by an
authorised Ottobock Service
Centre.
Error
Distance between bat­
tery charger and
receiver of the char­
ging unit on the pros­
thesis too great. If the
distance is more than
2 mm, the prosthesis
cannot be charged.
perature
Excessively high or
low temperature of the
prosthesis
The prosthesis is not
being charged
Distance between bat­
tery charger and
receiver of the char­
ging unit too great.
The battery charger is
functional but not con­
nected to the receiver,
or the distance
between the battery
charger and receiver
of the charging unit is
too great.
Resolution
Reduce distance
between battery char­
ger and receiver of the
charging unit.
Check whether the
specified ambient con­
ditions for charging the
battery are met (see
Page 95).
Connection may be
improved by reducing
the distance between
the charger and receiv­
er of the charging unit.
Connect the battery
charger or reduce the
distance between the
charger and the receiv­
er of the charging unit
on the prosthesis.
101

Advertisement

Chapters

Table of Contents
loading

This manual is also suitable for:

3b5-x23b5-x2-st

Table of Contents