System Connection; Readhead Only (No Interface) - Renishaw ATOM Installation Manual

Rotary encoder system
Hide thumbs Also See for ATOM:
Table of Contents

Advertisement

System connection

Readhead only
(no interface)
The ATOM readhead is available in several variants:
Cable variant with a 15 way D-type connector
Cable variant with interboard connector
FPC variant
None of these readhead variants has an integral calibration (CAL) button. Provision should be made
in the customers' electronics for momentarily connecting the CAL line to 0 V to initiate the calibration
routine, switch AGC on /off or restore factory defaults. See page
Calibration is an essential part of system set-up that optimises the incremental signals and phases the
reference mark. See page
34
for information on the calibration routine.
16
ATOM rotary encoder system
42
for pin-out information.
For cable variant with interboard connector
Ensure that the interboard connector is inserted into the connector on the customers' electronics
NOTE: Care should be taken to ensure correct orientation
Use a metal clamp around the cable ferrule to ground the readhead cable and ensure
continuity of the shields
Provide appropriate strain relief
Ensure suitable clamping to retain interboard connector to the mating connector
JST 12-way, single row 12SUR-32S.
Hirose
Interboard connector
JST mating connector: P-CN10-0051
14-way, dual row
(SM12B-SURS-TF).
DF23C-14DP-0.5 V
Hirose mating connector:
P-CN03-0024
(DF23-14DS-0.5 V)
1
13
2
14
Interboard connector

Advertisement

Table of Contents
loading

Table of Contents