might not offer adequate protection to radio-frequency communication
services. The user might need to take mitigation measures, such as relocating
or re-orienting the equipment.
Table 51: Guidance and manufacturer's declaration on electromagnetic
immunity
Immunity
IEC 60601-1-2
Test
Electrostatic
± 8 kV contact
discharge
± 2 kV, ± 4 kV,
(ESD)
± 8 kV, ± 15 kV
IEC 61000-4-2
air
Radiated RF
3 V/m
EM fields
from 80 MHz to
IEC 61000-4-3
2.7 GHz
Electrical fast
± 2 kV for power
transient/
supply lines
burst
± 1 kV for input/
IEC 61000-4-4
output lines
(> 3 m)
Surge
± 0.5 kV, ± 1 kV
line(s) to line(s)
IEC 61000-4-5
± 0.5 kV, ± 1 kV,
± 2 kV line(s) to
earth
Conducted
3 V
disturbances
from 150 kHz to
induced by RF
80 Mhz
fields
IEC 61000-4-6
Compliance
Test Level
± 8 kV contact
± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV
air
3 V/m
from 80 MHz to
2.7 GHz
± 2 kV for power
supply lines
± 1 kV for input/
output lines
(> 3 m)
± 0.5 kV, ± 1 kV
line(s) to line(s)
± 0.5 kV, ± 1 kV,
± 2 kV line(s) to
earth
3 V
from 150 kHz to
80 Mhz
DR 100s | Product Information | 193
Electromagnetic en-
Level
Floors should consist
of wood, concrete or
ceramic tiles. The rel-
ative humidity must
be at least 30%, if the
floor is made of syn-
thetic material.
Portable RF commu-
nications equipment
(including peripher-
als such as antenna
cables and external
antennas) should be
used no closer than
30 cm to any part of
the equipment, in-
cluding cables speci-
fied by manufacturer.
The quality of the
voltage supplied
should correspond to
a typical commercial
or clinical environ-
ment.
The quality of the
voltage supplied
should correspond to
that of a typical com-
mercial or clinical en-
vironment.
Portable RF commu-
nications equipment
(including peripher-
als such as antenna
cables and external
antennas) should be
0411D EN 20220627 1040
vironment