AGFA DR 100s User Manual page 180

Table of Contents

Advertisement

180 | DR 100s | Product Information
Table 34: Maximum settable mAs
focus
exposure
time
minimum
mAs
kV
40
50
60
70
80
90
100
110
120
130
140
150
Table 35: Generator
Model
Maximum power (100 kV - 400 mA)
Maximum voltage to the tube
Maximum current to the tube
Ripple to the maximum power
Rise time to the max. power
Thermal safety
Oil volume compensation lung
0411D EN 20220627 1040
small
small
1 s
3 s
0.1
0.1
maximum
maximum
mAs
mAs
200
280
200
280
160
250
160
250
125
180
125
180
125
180
100
160
100
160
80
125
80
125
80
125
large
large
1 s
3 s
0.63
0.63
maximum
maximum
mAs
mAs
320
500
320
500
320
500
320
500
250
360
250
360
200
280
200
280
160
250
160
250
125
250
125
250
HF1 G/29
40 kW
150 kV
450 mA
< 2%
< 1 ms
60 °C ± 5° C
3
0.20 dm

Advertisement

Table of Contents
loading

This manual is also suitable for:

6013/100

Table of Contents