Chapter 6: Testing Film Samples

This chapter contains information on how to test flat film samples. Suggestions on how to maximize the
accuracy of your results and procedures describing how to perform the test are discussed.
Read this chapter to learn about:

Testing Suggestions

Preparing for a Film Test
Sample Size
Using Masks
Sample Orientation
Preparing the Samples for Testing
Filling the Humidifier
Filling the Humidifier during a Test
Conducting a Film Test
Testing Suggestions
The following information will assist you in the day to day operation of your instrument.
Test duplicate samples whenever possible.
Use good samples. Pinholes and creases can contribute to false readings.
Use caution when recording data for a new or unknown specimen. Some materials approach
equilibrium slowly. When in doubt repeat the test.
If the test is to be run under precise RH conditions, make sure that the humidifier has
sufficient water in it.
If no testing will be performed for a significant period (overnight or weekend) no special
precautions are needed. The Gas Saver function will automatically be activated (when
enabled) whenever both test cells are in the "Idle" state.
Turning off the power and gas supply to instrument is not recommended. If the instrument is
shutdown you will need to wait for it to outgas before testing can resume.
MOCON, Inc.
Revision G
Testing Film Samples
6-1