(17) [OP1]
Contents:
Format:
Parameter:
Return value format:
Description & others:
(18) [SCV]
Contents:
Format:
Parameter:
Return value format:
Description & others:
(19) [SCV]
Contents:
Format:
Parameter:
Return value format:
Description & others:
Set analog output (A/D output). (Relevant to panel operation F-06.)
OP1020, ±xxxxxx,±yyyyyy<CR>
±xxxxxx,±yyyyyy
• Both indicated values include decimal point.
If the indicated value has no decimal point, it is filled
with '0' from the head.
ok<CR><LF>
Corresponding to '/M' command.
Set comparator
SCVx, ±yyyyyy<CR>
x, ±yyyyyy
ok<CR><LF>
Corresponding to '/M'command.
Set hysteresis. (Relevant to panel operation F-10.)
SCV0,xxxxxx<CR>
xxxxxx
Indicates the indicated value with a decimal point. If the indicated
value has no decimal point, it is filled with '0' from the head.
ok<CR><LF>
Corresponding to '/M' command.
44
Indicated value at 10 volts output
Indicated value at 0 volt output.
Comparative value
Indicated value with a decimal point
If the indicated value has no decimal point, it is filled
with '0' from the head.
Select the comparator.
1: Comparator H
2: Comparator L
3: Spare
4: Spare