KYOWA WGA-670B-0 Instruction Manual page 18

Instrumentation amplifier
Table of Contents

Advertisement

Function
Function Name
Selecting Mode
F-02
Actual Load
Calibration
F-03
Sensitivity
Registering
Calibration
Key operation & WGA-670B
Movement in Function Setting Mode
●The indication '00628' or '11111' is a
key code to prevent calibration value
from carelessly changing due to
malfunction.
Set transducer rated output (Unit: mV/V)
and press the SET key.
Next, set the required value to be
indicated and press the SET key.
Then, set a decimal point position and
press the SET key.
•Key code enable setting
'00000' is indicated.
Here, set the indication to read '00628' or
'11111' and press the SET key. Then,
calibration value is updated.
The key code is enabled. Therefore, when
the calibration value is updated the next
time, the above operation is required. To
disable the key code, use the function
'F-12.'
•Key code disable setting
'00000' is indicated. Press the SET key
as it is. Then, the calibration value is
updated.
The key code is disabled. The calibration
value is updated even if no key code is
entered.
●The indication '00628' or '11111' is a
key code to prevent calibration value
from carelessly changing due to
malfunction.
●Setting available range is '10000' or
less per 0.1 mV/V.
13
Functional
Description
•A function that
conducts calibra-
tion by registering
the rated output
and rated capacity
of the transducer
without applying
the actual load.
•Factory default is
'4000' when 2.000
mV/V is entered.
•For details, see
"4-2-4."