(11) [EIC]
Contents:
Format:
Parameter:
Return value format:
Description & others:
(12) [EID]
Contents:
Format:
Parameter:
Return value format:
Description & others:
(13) [SMF]
Contents:
Format:
Parameter:
Return value format:
Description & others:
Conducts sensitivity registering calibration.
(Relevant to panel operation F-03)
EIC0x.xxx, ±yyyyyy<CR>
0x.xxx, ±yyyyyy
ok<CR><LF>
When error occurs, E-00<CR><LF>
Saves the results of this command.
Automatically conducts sensitivity registering calibration with TEDS
information.
EID<CR>
None
ok<CR><LF>
When error occurs, E-03<CR><LF>
Among TEDS information, rated capacity, rated output, and recommended
excitation voltage are read to set sensitivity registration and excitation
voltage.
Set the hold mode. (Relevant to panel operation F-04.)
SMFa, b,c,x.xx,yy.yy<CR>
a, b, c,x.xx,yy.yy
Comparison mode
Display mode
Hold mode
ok<CR><LF>
Corresponding to '/M' command
Indicates the indicated value with a decimal point.
If the indicated value has no decimal point, it is
filled with '0' from the head.
Sensor rated output value (mV/V)
Specified time (sec)
Delay time (sec)
0: Normally compare 1: Compare at hold mode
0: Normally display
0: Tracking
1: Arbitrary point hold
2: Peak hold
3: Block specified peak hold
4: Time specified peak hold