Instrumentation amplifier
Table of Contents

Advertisement

Function
Function Name
Selecting Mode
F-13
BV Select
F-14
TEDS ON/OFF
Key Operation & WGA-670B
Movement in Function Setting Mode
BV set value
0: 10 V
1: 2 V
Select either of the two and press the SET
key.
Indicates the following items (1) and (2) .
(1) Sensitivity registration with power ON
0: OFF
1: ON
(2) Digital ZERO operation with power ON
0: OFF
1: ON
Select and change numerics in items (1) and
(2) and press the SET key.
(Enabled only for WGA-670B-7.)
18
Functional
Description
•A function that
selects bridge
exciter voltage.
•It is set to 1(2 V)
prior to
shipment.
•A function
whether or not
sensitivity
registration or
digital Zero is
conducted by
reading TEDS
information with
the power ON.
• Factory default is
'0-0.'